Quality assurance in proton beam therapy using a plastic scintillator and a commercially available digital camera.
Journal of Applied Clinical Medical Physics
, 18
(5)
pp. 210-219.
10.1002/acm2.12143.
Preview |
Text
Almurayshid_et_al-2017-Journal_of_Applied_Clinical_Medical_Physics.pdf - Accepted Version Download (627kB) | Preview |
Abstract
PURPOSE: In this article, we evaluate a plastic scintillation detector system for quality assurance in proton therapy using a BC-408 plastic scintillator, a commercial camera, and a computer. METHODS: The basic characteristics of the system were assessed in a series of proton irradiations. The reproducibility and response to changes of dose, dose-rate, and proton energy were determined. Photographs of the scintillation light distributions were acquired, and compared with Geant4 Monte Carlo simulations and with depth-dose curves measured with an ionization chamber. A quenching effect was observed at the Bragg peak of the 60 MeV proton beam where less light was produced than expected. We developed an approach using Birks equation to correct for this quenching. We simulated the linear energy transfer (LET) as a function of depth in Geant4 and found Birks constant by comparing the calculated LET and measured scintillation light distribution. We then used the derived value of Birks constant to correct the measured scintillation light distribution for quenching using Geant4. RESULTS: The corrected light output from the scintillator increased linearly with dose. The system is stable and offers short-term reproducibility to within 0.80%. No dose rate dependency was observed in this work. CONCLUSIONS: This approach offers an effective way to correct for quenching, and could provide a method for rapid, convenient, routine quality assurance for clinical proton beams. Furthermore, the system has the advantage of providing 2D visualization of individual radiation fields, with potential application for quality assurance of complex, time-varying fields.
Type: | Article |
---|---|
Title: | Quality assurance in proton beam therapy using a plastic scintillator and a commercially available digital camera |
Location: | United States |
Open access status: | An open access version is available from UCL Discovery |
DOI: | 10.1002/acm2.12143 |
Publisher version: | http://dx.doi.org/10.1002/acm2.12143 |
Language: | English |
Additional information: | This is an open access article under the terms of the Creative Commons Attribution License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited. © 2017 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine |
Keywords: | Proton therapy, quality assurance, radiotherapy |
UCL classification: | UCL UCL > Provost and Vice Provost Offices > UCL BEAMS UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science > Dept of Med Phys and Biomedical Eng |
URI: | https://discovery.ucl.ac.uk/id/eprint/1570219 |




Archive Staff Only
![]() |
View Item |