The PTW microSilicon diode: Performance in small 6 and 15 MV photon fields and utility of density compensation

Georgiou, G; Kumar, S; Würfel, JU; Gilmore, M; Underwood, TSA; Rowbottom, CG; Fenwick, JD; (2021) The PTW microSilicon diode: Performance in small 6 and 15 MV photon fields and utility of density compensation. Medical Physics , 48 (12) pp. 8062-8074. 10.1002/mp.15329. Green open access

[thumbnail of Fenwick_The PTW microSilicon diode- Performance in small 6 and 15 MV photon fields and utility of density compensation_AAM.pdf]
Preview
Text
Fenwick_The PTW microSilicon diode- Performance in small 6 and 15 MV photon fields and utility of density compensation_AAM.pdf

Download (691kB) | Preview

Abstract

PURPOSE: We have experimentally and computationally characterized the PTW microSilicon 60023-type diode's performance in 6 and 15 MV photon fields ≥5 × 5 mm2 projected to isocenter. We tested the detector on- and off-axis at 5 and 15 cm depths in water, and investigated whether its response could be improved by including within it a thin airgap. METHODS: Experimentally, detector readings were taken in fields generated by a Varian TrueBeam linac and compared with doses-to-water measured using Gafchromic film and ionization chambers. An unmodified 60023-type diode was tested along with detectors modified to include 0.6, 0.8, and 1.0 mm thick airgaps. Computationally, doses absorbed by water and detectors’ sensitive volumes were calculated using the EGSnrc/BEAMnrc Monte Carlo radiation transport code. Detector response was characterized using K^{fdin, 4cm}_{Qclin, 4cm}, a factor that corrects for differences in the ratio of dose-to-water to detector reading between small fields and the reference condition, in this study 5 cm deep on-axis in a 4 × 4 cm2 field. RESULTS: The greatest errors in measurements of small field doses made using uncorrected readings from the unmodified 60023-type detector were over-responses of 2.6% ± 0.5% and 5.3% ± 2.0% determined computationally and experimentally, relative to the reading-per-dose in the reference field. Corresponding largest errors for the earlier 60017-type detector were 11.9% ± 0.6% and 11.7% ± 1.4% over-responses. Adding even the thinnest, 0.6 mm, airgap to the 60023-type detector over-corrected it, leading to under-responses of up to 4.8% ± 0.6% and 5.0% ± 1.8% determined computationally and experimentally. Further, Monte Carlo calculations indicate that a detector with a 0.3 mm airgap would read correctly to within 1.3% on-axis. The ratio of doses at 15 and 5 cm depths in water in a 6 MV 4 × 4 cm2 field was measured more accurately using the unmodified 60023-type detector than using the 60017-type detector, and was within 0.3% of the ratio measured using an ion chamber. The 60023-type diode's sensitivity also varied negligibly as dose-rate was reduced from 13 to 4 Gy min–1 by decreasing the linac pulse repetition frequency, whereas the sensitivity of the 60017-type detector fell by 1.5%. CONCLUSIONS: The 60023-type detector performed well in small fields across a wide range of beam energies, field sizes, depths, and off-axis positions. Its response can potentially be further improved by adding a thin, 0.3 mm, airgap.

Type: Article
Title: The PTW microSilicon diode: Performance in small 6 and 15 MV photon fields and utility of density compensation
Location: United States
Open access status: An open access version is available from UCL Discovery
DOI: 10.1002/mp.15329
Publisher version: https://doi.org/10.1002/mp.15329
Language: English
Additional information: This version is the author accepted manuscript. For information on re-use, please refer to the publisher's terms and conditions.
Keywords: density compensation, diode, dose-rate, microSilicon, small field, Monte Carlo Method, Particle Accelerators, Photons, Radiometry, Water
UCL classification: UCL
UCL > Provost and Vice Provost Offices > UCL BEAMS
UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science
UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science > Dept of Med Phys and Biomedical Eng
URI: https://discovery.ucl.ac.uk/id/eprint/10161418
Downloads since deposit
190Downloads
Download activity - last month
Download activity - last 12 months
Downloads by country - last 12 months

Archive Staff Only

View Item View Item