Jaubert, O;
Montalt-Tordera, J;
Knight, D;
Coghlan, GJ;
Arridge, S;
Steeden, JA;
Muthurangu, V;
(2021)
Real-time deep artifact suppression using recurrent U-Nets for low-latency cardiac MRI.
Magnetic Resonance in Medicine
10.1002/mrm.28834.
(In press).
Preview |
Text
Jaubert_mrm.28834.pdf - Published Version Download (3MB) | Preview |
Abstract
PURPOSE: Real-time low latency MRI is performed to guide various cardiac interventions. Real-time acquisitions often require iterative image reconstruction strategies, which lead to long reconstruction times. In this study, we aim to reconstruct highly undersampled radial real-time data with low latency using deep learning. METHODS: A 2D U-Net with convolutional long short-term memory layers is proposed to exploit spatial and preceding temporal information to reconstruct highly accelerated tiny golden radial data with low latency. The network was trained using a dataset of breath-hold CINE data (including 770 time series from 7 different orientations). Synthetic paired data were created by retrospectively undersampling the magnitude images, and the network was trained to recover the target images. In the spirit of interventional imaging, the network was trained and tested for varying acceleration rates and orientations. Data were prospectively acquired and reconstructed in real time in 1 healthy subject interactively and in 3 patients who underwent catheterization. Images were visually compared to sliding window and compressed sensing reconstructions and a conventional Cartesian real-time sequence. RESULTS: The proposed network generalized well to different acceleration rates and unseen orientations for all considered metrics in simulated data (less than 4% reduction in structural similarity index compared to similar acceleration and orientation-specific networks). The proposed reconstruction was demonstrated interactively, successfully depicting catheters in vivo with low latency (39 ms, including 19 ms for deep artifact suppression) and an image quality comparing favorably to other reconstructions. CONCLUSION: Deep artifact suppression was successfully demonstrated in the time-critical application of non-Cartesian real-time interventional cardiac MR.
Type: | Article |
---|---|
Title: | Real-time deep artifact suppression using recurrent U-Nets for low-latency cardiac MRI |
Location: | United States |
Open access status: | An open access version is available from UCL Discovery |
DOI: | 10.1002/mrm.28834 |
Publisher version: | https://doi.org/10.1002/mrm.28834 |
Language: | English |
Additional information: | This work is licensed under a Creative Commons Attribution 4.0 International License. The images or other third party material in this article are included in the Creative Commons license, unless indicated otherwise in the credit line; if the material is not included under the Creative Commons license, users will need to obtain permission from the license holder to reproduce the material. To view a copy of this license, visit http://creativecommons.org/licenses/by/4.0/ |
Keywords: | cardiac MRI, deep learning, image reconstruction, interventional, real-time |
UCL classification: | UCL UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Population Health Sciences > Institute of Cardiovascular Science UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Population Health Sciences > Institute of Cardiovascular Science > Childrens Cardiovascular Disease UCL > Provost and Vice Provost Offices > UCL BEAMS UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science > Dept of Computer Science |