Qi, J;
Elson, DS;
Stoyanov, D;
(2019)
Eigenvalue calibration method for 3 × 3 Mueller polarimeters.
Optics Letters
, 44
(9)
pp. 2362-2365.
10.1364/OL.44.002362.
Preview |
Text
ol-44-9-2362.pdf - Published Version Download (1MB) | Preview |
Abstract
3×3 Mueller polarimetry has shown potential for tissue characterization applications, however, calibration has not been fully addressed. We demonstrate a 3×3 Mueller polarimeter eigenvalue calibration method, inspired by those for full Mueller polarimeters. We also investigate the optimal combination of calibration measurements. Our method does not rely on modeling the polarization state generator, polarization state analyzer, or precise knowledge of calibration sample properties or orientations. It is therefore easy to implement, and the experimental results of a linear polarizer test sample, as well as a biological specimen, are presented.
Type: | Article |
---|---|
Title: | Eigenvalue calibration method for 3 × 3 Mueller polarimeters |
Location: | United States |
Open access status: | An open access version is available from UCL Discovery |
DOI: | 10.1364/OL.44.002362 |
Publisher version: | https://doi.org/10.1364/OL.44.002362 |
Language: | English |
Additional information: | Published by The Optical Society under the terms of the Creative Commons Attribution 4.0 License (http://creativecommons.org/licenses/by/4.0/). Further distribution of this work must maintain attribution to the author(s) and the published article's title, journal citation, and DOI. |
UCL classification: | UCL UCL > Provost and Vice Provost Offices > UCL BEAMS UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science > Dept of Computer Science |
URI: | https://discovery.ucl.ac.uk/id/eprint/10075024 |



1. | ![]() | 3 |
2. | ![]() | 1 |
3. | ![]() | 1 |
Archive Staff Only
![]() |
View Item |