Lee, Y;
Callaghan, MF;
Acosta-Cabronero, J;
Lutti, A;
Nagy, Z;
(2019)
Establishing intra- and inter-vendor reproducibility of T₁ relaxation time measurements with 3T MRI.
Magnetic Resonance in Medicine
, 81
(1)
pp. 454-465.
10.1002/mrm.27421.
Preview |
Text
Callaghan_MRM_Rev(final_with_References).pdf - Accepted Version Download (496kB) | Preview |
Abstract
PURPOSE: Parametric imaging methods (e.g., T1 relaxation time mapping) have been shown to be more reproducible across time and vendors than weighted (e.g., T1 -weighted) images. The purpose of this work was to more extensively evaluate the validity of this assertion. METHODS: Seven volunteers underwent twice-repeated acquisitions of variable flip-angle T1 mapping, including B1+ calibration, on a 3T Philips Achieva and 3T Siemens Trio scanner. Intra-scanner and inter-vendor T1 variability were calculated. To determine T1 reproducibility levels in longitudinal settings, or after changing hardware or software, four additional data sets were acquired from two of the participants; one participant was scanned on a different 3T Siemens Trio scanner and another on the same 3T Philips Achieva scanner but after a software upgrade. RESULTS: Intra-scanner variability of voxel-wise T1 values was consistent between the two vendors, averaging 0.7/0.7/1.3/1.4% in white matter/cortical gray matter/subcortical gray matter/cerebellum, respectively. We observed, however, a systematic bias between the two vendors of https://doi.org/10.0/7.8/8.6/10.0%, respectively. The T1 bias across two scanners of the same model was greater than intra-scanner variability, although still only at 1.4/1.0/1.9/2.3%, respectively. A greater bias was identified for data sets acquired before/after software upgrade in white matter/cortical gray matter (3.6/2.7%) whereas variability in subcortical gray matter/cerebellum was comparable (1.7/1.9%). CONCLUSION: We established intra- and inter-vendor reproducibility levels for a widely used T1 mapping protocol. We anticipate that these results will guide the design of multi-center studies, particularly those encompassing multiple vendors. Furthermore, this baseline level of reproducibility should be established or surpassed during the piloting phase of such studies.
Type: | Article |
---|---|
Title: | Establishing intra- and inter-vendor reproducibility of T₁ relaxation time measurements with 3T MRI |
Location: | United States |
Open access status: | An open access version is available from UCL Discovery |
DOI: | 10.1002/mrm.27421 |
Publisher version: | https://doi.org/10.1002/mrm.27421 |
Language: | English |
Additional information: | This version is the author accepted manuscript. For information on re-use, please refer to the publisher’s terms and conditions. |
Keywords: | 3T, T1 relaxation, bias, multi-vendor, parametric imaging, reproducibility |
UCL classification: | UCL UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Brain Sciences UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Brain Sciences > UCL Queen Square Institute of Neurology UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Brain Sciences > UCL Queen Square Institute of Neurology > Imaging Neuroscience |