Bozkurt, S;
Preston-Maher, GL;
Torii, R;
Burriesci, G;
(2017)
Design, Analysis and Testing of a Novel Mitral Valve for Transcatheter Implantation.
Annals of Biomedical Engineering
, 45
(8)
pp. 1852-1864.
10.1007/s10439-017-1828-2.
Preview |
Text
art%3A10.1007%2Fs10439-017-1828-2.pdf - Published Version Download (6MB) | Preview |
Abstract
Mitral regurgitation is a common mitral valve dysfunction which may lead to heart failure. Because of the rapid aging of the population, conventional surgical repair and replacement of the pathological valve are often unsuitable for about half of symptomatic patients, who are judged high-risk. Transcatheter valve implantation could represent an effective solution. However, currently available aortic valve devices are inapt for the mitral position. This paper presents the design, development and hydrodynamic assessment of a novel bi-leaflet mitral valve suitable for transcatheter implantation. The device consists of two leaflets and a sealing component made from bovine pericardium, supported by a self-expanding wireframe made from superelastic NiTi alloy. A parametric design procedure based on numerical simulations was implemented to identify design parameters providing acceptable stress levels and maximum coaptation area for the leaflets. The wireframe was designed to host the leaflets and was optimised numerically to minimise the stresses for crimping in an 8 mm sheath for percutaneous delivery. Prototypes were built and their hydrodynamic performances were tested on a cardiac pulse duplicator, in compliance with the ISO5840-3:2013 standard. The numerical results and hydrodynamic tests show the feasibility of the device to be adopted as a transcatheter valve implant for treating mitral regurgitation.
Type: | Article |
---|---|
Title: | Design, Analysis and Testing of a Novel Mitral Valve for Transcatheter Implantation |
Location: | United States |
Open access status: | An open access version is available from UCL Discovery |
DOI: | 10.1007/s10439-017-1828-2 |
Publisher version: | http://doi.org/10.1007/s10439-017-1828-2 |
Language: | English |
Additional information: | © The Author(s) 2017. This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. |
Keywords: | Bioprosthetic bi-leaflet valve, Heart valve assessment, Heart valve development, Mitral valve, Transcatheter mitral valve implantation (TMVI) |
UCL classification: | UCL UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Population Health Sciences > Institute of Cardiovascular Science UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Population Health Sciences > Institute of Cardiovascular Science > Childrens Cardiovascular Disease UCL > Provost and Vice Provost Offices > UCL BEAMS UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science > Dept of Mechanical Engineering |
URI: | https://discovery.ucl.ac.uk/id/eprint/1547544 |




Archive Staff Only
![]() |
View Item |