Lourenco, A;
Thomas, R;
Homer, M;
Bouchard, H;
Rossomme, S;
Renaud, J;
Kanai, T;
... Palmans, H; + view all
(2017)
Fluence correction factor for graphite calorimetry in a clinical high-energy carbon-ion beam.
Physics in Medicine and Biology
, 62
N134-N146.
10.1088/1361-6560/aa6147.
Preview |
Text
Lourenco+et+al_2017_Phys._Med._Biol._10.1088_1361-6560_aa6147.pdf - Accepted Version Download (563kB) | Preview |
Abstract
The aim of this work is to develop and adapt a formalism to determine absorbed dose to water from graphite calorimetry measurements in carbon-ion beams. Fluence correction factors, <i>k</i><sub>fl</sub>, needed when using a graphite calorimeter to derive dose to water, were determined in a clinical high-energy carbon-ion beam. Measurements were performed in a 290 MeV/n carbon-ion beam with a field size of 11 x 11 cm<sup>2</sup>, without modulation. In order to sample the beam, a plane-parallel Roos ionization chamber was chosen for its small collecting volume in comparison with the field size. Experimental information on fluence corrections was obtained from depth-dose measurements in water. This procedure was repeated with graphite plates in front of the water phantom. Fluence corrections were also obtained with Monte Carlo simulations through the implementation of three methods based on (i) the fluence distributions differential in energy, (ii) a ratio of calculated doses in water and graphite at equivalent depths and (iii) simulations of the experimental setup. The <i>k</i><sub>fl</sub> term increased in depth from 1.00 at the entrance toward 1.02 at a depth near the Bragg peak, and the average difference between experimental and numerical simulations was about 0.13%. Compared to proton beams, there was no reduction of the <i>k</i><sub>fl</sub> due to alpha particles because the secondary particle spectrum is dominated by projectile fragmentation. By developing a practical dose conversion technique, this work contributes to improving the determination of absolute dose to water from graphite calorimetry in carbon-ion beams.
Type: | Article |
---|---|
Title: | Fluence correction factor for graphite calorimetry in a clinical high-energy carbon-ion beam |
Location: | England |
Open access status: | An open access version is available from UCL Discovery |
DOI: | 10.1088/1361-6560/aa6147 |
Publisher version: | http://dx.doi.org/10.1088/1361-6560/aa6147 |
Language: | English |
Additional information: | As the Version of Record of this article is going to be/has been published on a gold open access basis under a CC BY 3.0 licence, this Accepted Manuscript is available for reuse under a CC BY 3.0 licence immediately. Although reasonable endeavours have been taken to obtain all necessary permissions from third parties to include their copyrighted content within this article, their full citation and copyright line may not be present in this Accepted Manuscript version. Before using any content from this article, please refer to the Version of Record on IOPscience once published for full citation and copyright details, as permission may be required. All third party content is fully copyright protected, and is not published on a gold open access basis under a CC BY licence, unless that is specifically stated in the figure caption in the Version of Record. |
Keywords: | absolute dosimetry, carbon-ion therapy, conversion factors, fluence corrections, graphite calorimetry |
UCL classification: | UCL UCL > Provost and Vice Provost Offices > UCL BEAMS UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science > Dept of Med Phys and Biomedical Eng |