Iatrogenic cardiac perforation due to pacemaker and defibrillator leads: a contemporary multicentre experience

Waddingham, Peter H; Elliott, James; Bates, Alexander; Bilham, James; Muthumala, Amal; Honarbakhsh, Shohreh; Ullah, Waqas; ... Chow, Anthony WC; + view all (2022) Iatrogenic cardiac perforation due to pacemaker and defibrillator leads: a contemporary multicentre experience. EP Europace , 24 (11) pp. 1824-1833. 10.1093/europace/euac105. Green open access

[thumbnail of Lead perforation Europace.pdf]
Preview
PDF
Lead perforation Europace.pdf - Published Version

Download (719kB) | Preview

Abstract

AIMS: To determine the incidence, clinical features, management, and outcomes of pacemaker (PM) and implantable cardioverter-defibrillator (ICD) lead cardiac perforation. Cardiac perforations due to PM and ICD leads are rare but serious complications. Clinical features vary widely and may cause diagnostic delay. Management strategies are non-guideline based due to paucity of data. METHODS AND RESULTS: A multicentre retrospective series including 3 UK cardiac tertiary centres from 2016 to 2020. Patient, device, and lead characteristics were obtained including 6-month outcomes. Seventy cases of perforation were identified from 10 631 procedures; perforation rate was 0.50% for local implants. Thirty-nine (56%) patients were female, mean ( ± standard deviation) age 74 ( ± 13.8) years. Left ventricular ejection fraction 51 ( ± 13.2) %. Median time to diagnosis was 9 (range: 0-989) days. Computed tomography (CT) diagnosed perforation with 97% sensitivity. Lead parameter abnormalities were present in 86% (whole cohort) and 98.6% for perforations diagnosed >24 h. Chest pain was the commonest symptom, present in 46%. The management strategy was percutaneous in 98.6% with complete procedural success in 98.6%. Pericardial effusion with tamponade was present in 17% and was associated with significantly increased mortality and major complications. Anticoagulation status was associated with tamponade by multivariate analysis (odds ratio 21.7, 95% confidence interval: 1.7-275.5, P = 0.018). CONCLUSIONS: Perforation was rare (0.50%) and managed successfully by a percutaneous strategy with good outcomes. Tamponade was associated with increased mortality and major complications. Anticoagulation status was an independent predictor of tamponade. Case complexity is highly variable and requires skilled operators with a multi-disciplinary approach to achieve good outcomes.

Type: Article
Title: Iatrogenic cardiac perforation due to pacemaker and defibrillator leads: a contemporary multicentre experience
Location: England
Open access status: An open access version is available from UCL Discovery
DOI: 10.1093/europace/euac105
Publisher version: http://dx.doi.org/10.1093/europace/euac105
Language: English
Additional information: This version is the version of record. For information on re-use, please refer to the publisher’s terms and conditions.
Keywords: Cardiac perforation, Defibrillator, Lead, Pacemaker, Pericardial effusion, Tamponade
UCL classification: UCL
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Population Health Sciences > Institute of Cardiovascular Science
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Population Health Sciences > Institute of Cardiovascular Science > Clinical Science
URI: https://discovery.ucl.ac.uk/id/eprint/10194612
Downloads since deposit
10Downloads
Download activity - last month
Download activity - last 12 months
Downloads by country - last 12 months

Archive Staff Only

View Item View Item