Havariyoun, G;
Massimi, L;
Hagen, C;
Endrizzi, M;
Olivo, A;
(2023)
Modulation Transfer Function (MTF) evaluation for x-ray phase imaging system employing attenuation masks.
Physics in Medicine and Biology
10.1088/1361-6560/acc927.
(In press).
Preview |
Text
Olivo_Havariyoun+et+al_2023_Phys._Med._Biol._10.1088_1361-6560_acc927.pdf Download (908kB) | Preview |
Abstract
OBJECTIVE: Attenuation masks can be used in x-ray imaging systems to increase their inherent spatial resolution and/or make them sensitive to phase effects, a typical example being Edge Illumination X-ray phase contrast imaging (EI-XPCI). This work investigates the performance of a mask-based system such as EI-XPCI in terms of Modulation Transfer Function (MTF), in the absence of phase effects. APPROACH: Pre-sampled MTF measurements, using an edge, were performed on the same system implemented without masks, with non-skipped masks and finally with skipped masks (i.e., masks in which apertures illuminate every other pixel row/column). Results are compared to simulations and finally images of a resolution bar pattern acquired with all the above setups are presented. MAIN RESULTS: Compared to the detector's inherent MTF, the non-skipped mask setup provides improved MTF results. In comparison to an ideal case where signal spill-out into neighbouring pixels is negligible, this improvement takes place only at specific frequencies of the MTF, dictated by the spatial repetition of the spill-out signal. This is limited with skipped masks, which indeed provide further MTF improvements over a larger frequency range. Experimental MTF measurements are supported through simulation and resolution bar pattern images. SIGNIFICANCE: This work has quantified the improvement in MTF due to the use of attenuation masks and lays the foundation for how acceptance and routine quality control tests will have to be modified when systems using masks are introduced in clinical practice and how MTF results will compare to those of conventional imaging systems.
Type: | Article |
---|---|
Title: | Modulation Transfer Function (MTF) evaluation for x-ray phase imaging system employing attenuation masks |
Open access status: | An open access version is available from UCL Discovery |
DOI: | 10.1088/1361-6560/acc927 |
Publisher version: | https://doi.org/10.1088/1361-6560/acc927 |
Language: | English |
Additional information: | As the Version of Record of this article is going to be / has been published on a gold open access basis under a CC BY 4.0 licence, this Accepted Manuscript is available for reuse under a CC BY 4.0 licence immediately. Everyone is permitted to use all or part of the original content in this article, provided that they adhere to all the terms of the licence https://creativecommons.org/licences/by/4.0 Although reasonable endeavours have been taken to obtain all necessary permissions from third parties to include their copyrighted content within this article, their full citation and copyright line may not be present in this Accepted Manuscript version. Before using any content from this article, please refer to the Version of Record on IOPscience once published for full citation and copyright details, as permissions may be required. All third party content is fully copyright protected and is not published on a gold open access basis under a CC BY licence, unless that is specifically stated in the figure caption in the Version of Record. |
UCL classification: | UCL UCL > Provost and Vice Provost Offices > UCL BEAMS UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science > Dept of Med Phys and Biomedical Eng |
URI: | https://discovery.ucl.ac.uk/id/eprint/10167573 |




Archive Staff Only
![]() |
View Item |