Kiersnowski, Oliver C;
Karsa, Anita;
Wastling, Stephen J;
Thornton, John S;
Shmueli, Karin;
(2022)
Investigating the effect of oblique image acquisition on the accuracy of QSM and a robust tilt correction method.
Magnetic Resonance in Medicine
10.1002/mrm.29550.
(In press).
Preview |
Text
Investigating the effect of oblique image acquisition.pdf - Published Version Download (4MB) | Preview |
Abstract
PURPOSE: Quantitative susceptibility mapping (QSM) is used increasingly for clinical research where oblique image acquisition is commonplace, but its effects on QSM accuracy are not well understood. THEORY AND METHODS: The QSM processing pipeline involves defining the unit magnetic dipole kernel, which requires knowledge of the direction of the main magnetic field B ^ 0 $$ {\hat{\boldsymbol{B}}}_{\mathbf{0}} $$ with respect to the acquired image volume axes. The direction of B ^ 0 $$ {\hat{\boldsymbol{B}}}_{\mathbf{0}} $$ is dependent on the axis and angle of rotation in oblique acquisition. Using both a numerical brain phantom and in vivo acquisitions in 5 healthy volunteers, we analyzed the effects of oblique acquisition on magnetic susceptibility maps. We compared three tilt-correction schemes at each step in the QSM pipeline: phase unwrapping, background field removal and susceptibility calculation, using the RMS error and QSM-tuned structural similarity index. RESULTS: Rotation of wrapped phase images gave severe artifacts. Background field removal with projection onto dipole fields gave the most accurate susceptibilities when the field map was first rotated into alignment with B ^ 0 $$ {\hat{\boldsymbol{B}}}_{\mathbf{0}} $$ . Laplacian boundary value and variable-kernel sophisticated harmonic artifact reduction for phase data background field removal methods gave accurate results without tilt correction. For susceptibility calculation, thresholded k-space division, iterative Tikhonov regularization, and weighted linear total variation regularization, all performed most accurately when local field maps were rotated into alignment with B ^ 0 $$ {\hat{\boldsymbol{B}}}_{\mathbf{0}} $$ before susceptibility calculation. CONCLUSION: For accurate QSM, oblique acquisition must be taken into account. Rotation of images into alignment with B ^ 0 $$ {\hat{\boldsymbol{B}}}_{\mathbf{0}} $$ should be carried out after phase unwrapping and before background-field removal. We provide open-source tilt-correction code to incorporate easily into existing pipelines: https://github.com/o-snow/QSM_TiltCorrection.git.
Type: | Article |
---|---|
Title: | Investigating the effect of oblique image acquisition on the accuracy of QSM and a robust tilt correction method |
Location: | United States |
Open access status: | An open access version is available from UCL Discovery |
DOI: | 10.1002/mrm.29550 |
Publisher version: | https://doi.org/10.1002/mrm.29550 |
Language: | English |
Additional information: | © 2022 The Authors. Magnetic Resonance in Medicine published by Wiley Periodicals LLC on behalf of International Society for Magnetic Resonance in Medicine. This is an open access article under the terms of the Creative Commons Attribution License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited. |
Keywords: | QSM, QSM accuracy, electromagnetic tissue properties, oblique acquisition, quantitative susceptibility mapping, tilted slices |
UCL classification: | UCL UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences UCL > Provost and Vice Provost Offices > UCL BEAMS UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Brain Sciences UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Brain Sciences > UCL Queen Square Institute of Neurology UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science > Dept of Med Phys and Biomedical Eng |
URI: | https://discovery.ucl.ac.uk/id/eprint/10161675 |
Archive Staff Only
View Item |