Ribeiro, D;
Hallett, W;
Howes, O;
McCutcheon, R;
Nour, MM;
Tavares, AAS;
(2022)
Assessing the impact of different penalty factors of the Bayesian reconstruction algorithm Q.Clear on in vivo low count kinetic analysis of [11C]PHNO brain PET-MR studies.
EJNMMI Research
, 12
, Article 11. 10.1186/s13550-022-00883-1.
Preview |
Text
s13550-022-00883-1.pdf - Published Version Download (3MB) | Preview |
Abstract
Introduction: Q.Clear is a Bayesian penalised likelihood (BPL) reconstruction algorithm available on General Electric (GE) Positron Emission Tomography (PET)-Computed Tomography (CT) and PET-Magnetic Resonance (MR) scanners. This algorithm is regulated by a β value which acts as a noise penalisation factor and yields improvements in signal to noise ratio (SNR) in clinical scans, and in contrast recovery and spatial resolution in phantom studies. However, its performance in human brain imaging studies remains to be evaluated in depth. This pilot study aims to investigate the impact of Q.Clear reconstruction methods using different β value versus ordered subset expectation maximization (OSEM) on brain kinetic modelling analysis of low count brain images acquired in the PET-MR. Methods: Six [11C]PHNO PET-MR brain datasets were reconstructed with Q.Clear with β100–1000 (in increments of 100) and OSEM. The binding potential relative to non-displaceable volume (BPND) were obtained for the Substantia Nigra (SN), Striatum (St), Globus Pallidus (GP), Thalamus (Th), Caudate (Cd) and Putamen (Pt), using the MIAKAT™ software. Intraclass correlation coefficients (ICC), repeatability coefficients (RC), coefficients of variation (CV) and bias from Bland–Altman plots were reported. Statistical analysis was conducted using a 2-way ANOVA model with correction for multiple comparisons. Results: When comparing a standard OSEM reconstruction of 6 iterations/16 subsets and 5 mm filter with Q.Clear with different β values under low counts, the bias and RC were lower for Q.Clear with β100 for the SN (RC = 2.17), Th (RC = 0.08) and GP (RC = 0.22) and with β200 for the St (RC = 0.14), Cd (RC = 0.18)and Pt (RC = 0.10). The p-values in the 2-way ANOVA model corroborate these findings. ICC values obtained for Th, St, GP, Pt and Cd demonstrate good reliability (0.87, 0.99, 0.96, 0.99 and 0.96, respectively). For the SN, ICC values demonstrate poor reliability (0.43). Conclusion: BPND results obtained from quantitative low count brain PET studies using [11C]PHNO and reconstructed with Q.Clear with β < 400, which is the value used for clinical [18F]FDG whole-body studies, demonstrate the lowest bias versus the typical iterative reconstruction method OSEM.
Type: | Article |
---|---|
Title: | Assessing the impact of different penalty factors of the Bayesian reconstruction algorithm Q.Clear on in vivo low count kinetic analysis of [11C]PHNO brain PET-MR studies |
Open access status: | An open access version is available from UCL Discovery |
DOI: | 10.1186/s13550-022-00883-1 |
Publisher version: | https://doi.org/10.1186/s13550-022-00883-1 |
Language: | English |
Additional information: | This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. |
UCL classification: | UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Brain Sciences UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Brain Sciences > UCL Queen Square Institute of Neurology > Imaging Neuroscience UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences UCL UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Brain Sciences > UCL Queen Square Institute of Neurology |
URI: | https://discovery.ucl.ac.uk/id/eprint/10145368 |
Archive Staff Only
View Item |