Tadimalla, Sirisha;
Wilson, Daniel J;
Shelley, David;
Bainbridge, Gavin;
Saysell, Margaret;
Mendichovszky, Iosif A;
Graves, Martin J;
... Sourbron, Steven P; + view all
(2022)
Bias, Repeatability and Reproducibility of Liver T1 Mapping With Variable Flip Angles.
Journal of Magnetic Resonance Imaging
10.1002/jmri.28127.
(In press).
Preview |
Text
Magnetic Resonance Imaging - 2022.pdf - Published Version Download (1MB) | Preview |
Abstract
BACKGROUND: Three-dimensional variable flip angle (VFA) methods are commonly used for T1 mapping of the liver, but there is no data on the accuracy, repeatability, and reproducibility of this technique in this organ in a multivendor setting. PURPOSE: To measure bias, repeatability, and reproducibility of VFA T1 mapping in the liver. STUDY TYPE: Prospective observational. POPULATION: Eight healthy volunteers, four women, with no known liver disease. FIELD STRENGTH/SEQUENCE: 1.5-T and 3.0-T; three-dimensional steady-state spoiled gradient echo with VFAs; Look-Locker. ASSESSMENT: Traveling volunteers were scanned twice each (30 minutes to 3 months apart) on six MRI scanners from three vendors (GE Healthcare, Philips Medical Systems, and Siemens Healthineers) at two field strengths. The maximum period between the first and last scans among all volunteers was 9 months. Volunteers were instructed to abstain from alcohol intake for at least 72 hours prior to each scan and avoid high cholesterol foods on the day of the scan. STATISTICAL TESTS: Repeated measures ANOVA, Student t-test, Levene's test of variances, and 95% significance level. The percent error relative to literature liver T1 in healthy volunteers was used to assess bias. The relative error (RE) due to intrascanner and interscanner variation in T1 measurements was used to assess repeatability and reproducibility. RESULTS: The 95% confidence interval (CI) on the mean bias and mean repeatability RE of VFA T1 in the healthy liver was 34 ± 6% and 10 ± 3%, respectively. The 95% CI on the mean reproducibility RE at 1.5 T and 3.0 T was 29 ± 7% and 25 ± 4%, respectively. DATA CONCLUSION: Bias, repeatability, and reproducibility of VFA T1 mapping in the liver in a multivendor setting are similar to those reported for breast, prostate, and brain. LEVEL OF EVIDENCE: 1 TECHNICAL EFFICACY STAGE: 1.
Type: | Article |
---|---|
Title: | Bias, Repeatability and Reproducibility of Liver T1 Mapping With Variable Flip Angles |
Location: | United States |
Open access status: | An open access version is available from UCL Discovery |
DOI: | 10.1002/jmri.28127 |
Publisher version: | https://doi.org/10.1002/jmri.28127 |
Language: | English |
Additional information: | This is an open access article under the terms of the Creative Commons Attribution License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited. |
Keywords: | T1 mapping, VFA, bias, liver, precision, repeatability, reproducibility |
UCL classification: | UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science > Dept of Med Phys and Biomedical Eng UCL > Provost and Vice Provost Offices > UCL BEAMS UCL |
URI: | https://discovery.ucl.ac.uk/id/eprint/10144616 |