UCL Discovery
UCL home » Library Services » Electronic resources » UCL Discovery

Temporary device malfunction of an MR conditional cardiac resynchronization defibrillator when undergoing MRI without appropriate re-programming: a case report

Patel, B; Monkhouse, C; Manisty, C; Papageorgiou, N; (2021) Temporary device malfunction of an MR conditional cardiac resynchronization defibrillator when undergoing MRI without appropriate re-programming: a case report. European Heart Journal. Case reports , 5 (9) pp. 1-6. 10.1093/ehjcr/ytab198. Green open access

[thumbnail of ytab198.pdf]
Preview
Text
ytab198.pdf - Published Version

Download (525kB) | Preview

Abstract

Background Magnetic resonance (MR) imaging (MRI) for patients with implantable cardiac devices is becoming more routine, with the development of MR conditional devices allowing more patients access to the imaging they need. However, for this to be performed safely, strict protocols must be followed necessitating close collaboration between cardiology and radiology departments. We present a case where mandatory device re-programming of a cardiac resynchronization therapy defibrillator device into MRI mode was not performed pre-scan leading to temporary device dysfunction with no clinical consequences. Case summary A 72-year-old man presented to a device clinic for a routine device interrogation. An atrial tachycardia response episode was recorded at the same time as the patient reported having undergone an MRI scan at a local centre. The electrogram demonstrated temporary right ventricular loss of capture with standard output programming, and a short episode of oversensing on the atrial and ventricular channel which was not sustained for long enough to meet tachycardia detection. Discussion We demonstrate two potential electrophysiological effects of MRI on pacemakers, where the device had not been appropriately re-programmed pre-procedure. This illustrates that whilst MRI in patients with implantable cardiac devices is safe, strict protocols must be followed requiring robust multidisciplinary communication.

Type: Article
Title: Temporary device malfunction of an MR conditional cardiac resynchronization defibrillator when undergoing MRI without appropriate re-programming: a case report
Open access status: An open access version is available from UCL Discovery
DOI: 10.1093/ehjcr/ytab198
Publisher version: https://doi.org/10.1093/ehjcr/ytab198
Language: English
Additional information: This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/4.0/), which permits non-commercial re-use, distribution, and reproduction in any medium, provided the original work is properly cited. For commercial re-use, please contact journals.permissions@oup.com
UCL classification: UCL
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Population Health Sciences > Institute of Cardiovascular Science
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Population Health Sciences > Institute of Cardiovascular Science > Clinical Science
URI: https://discovery.ucl.ac.uk/id/eprint/10135261
Downloads since deposit
33Downloads
Download activity - last month
Download activity - last 12 months
Downloads by country - last 12 months