Technical Note: Practical implementation strategies of cycloidal computed tomography

Roche i Morgó, O; Vittoria, F; Endrizzi, M; Olivo, A; Hagen, CK; (2021) Technical Note: Practical implementation strategies of cycloidal computed tomography. Medical Physics 10.1002/mp.14821. (In press). Green open access

[thumbnail of mp.14821.pdf]
Preview
Text
mp.14821.pdf - Published Version

Download (1MB) | Preview

Abstract

Purpose: Cycloidal computed tomography is a novel imaging concept which combines a highly structured x-ray beam, offset lateral under-sampling, and mathematical data recovery to obtain high-resolution images efficiently and flexibly, even with relatively large source focal spots and detector pixels. The method reduces scanning time and, potentially, delivered dose compared to other sampling schemes. This study aims to present and discuss several implementation strategies for cycloidal computed tomography (CT) in order to increase its ease of use and facilitate uptake within the imaging community. Methods: The different implementation strategies presented are step-and-shoot, continuous unidirectional, continuous back-and-forth, and continuous pixel-wise scanning. In step-and-shoot scans the sample remains stationary while frames are acquired, whereas in all other cases the sample moves through the scanner continuously. The difference between the continuous approaches is the trajectory by which the sample moves within the field of view. Results: All four implementation strategies are compatible with a standard table-top x-ray setup. With the experimental setup applied here, step-and-shoot acquisitions yield the best spatial resolution (around 30 µm), but are the most time-consuming (1.4 h). Continuous unidirectional and back-and-forth images have resolution between 30 and 40 µm, and are faster (35 min). Continuous pixel-wise images are equally time-efficient, although technical challenges caused a small loss in image quality with a resolution of about 50 µm. Conclusion: The authors show that cycloidal CT can be implemented in a variety of ways with high quality results. They believe this posits cycloidal CT as a powerful imaging alternative to more time-consuming and less flexible methods in the field.

Type: Article
Title: Technical Note: Practical implementation strategies of cycloidal computed tomography
Open access status: An open access version is available from UCL Discovery
DOI: 10.1002/mp.14821
Publisher version: https://doi.org/10.1002/mp.14821
Language: English
Additional information: © 2021 The Authors. Medical Physics published by Wiley Periodicals LLC on behalf of American Association of Physicists in Medicine. This is an open access article under the terms of the Creative Commons Attribution-NonCommercial License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited and is not used for commercial purposes.
UCL classification: UCL
UCL > Provost and Vice Provost Offices > UCL BEAMS
UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science
UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science > Dept of Med Phys and Biomedical Eng
URI: https://discovery.ucl.ac.uk/id/eprint/10130913
Downloads since deposit
Loading...
47Downloads
Download activity - last month
Loading...
Download activity - last 12 months
Loading...
Downloads by country - last 12 months
Loading...

Archive Staff Only

View Item View Item