UCL Discovery
UCL home » Library Services » Electronic resources » UCL Discovery

Does diametrical clearance influence the wear of Pinnacle hip implants?

Bergiers, S; Hothi, H; Henckel, J; Eskelinen, A; Skinner, J; Hart, A; (2020) Does diametrical clearance influence the wear of Pinnacle hip implants? Bone & Joint Research , 9 (8) pp. 515-523. 10.1302/2046-3758.98.BJR-2019-0195.R2. Green open access

[thumbnail of Hart_Does diametrical clearance influence the wear of Pinnacle hip implants?_VoR.pdf]
Preview
Text
Hart_Does diametrical clearance influence the wear of Pinnacle hip implants?_VoR.pdf - Published Version

Download (2MB) | Preview

Abstract

Aims: The optimum clearance between the bearing surfaces of hip arthroplasties is unknown. Theoretically, to minimize wear, it is understood that clearances must be low enough to maintain optimal contact pressure and fluid film lubrication, while being large enough to allow lubricant recovery and reduce contact patch size. This study aimed to identify the relationship between diametrical clearance and volumetric wear, through the analysis of retrieved components. / Methods: A total of 81 metal-on-metal Pinnacle hips paired with 12/14 stems were included in this study. Geometrical analysis was performed on each component, using coordinate and roundness measuring machines. The relationship between their as-manufactured diametrical clearance and volumetric wear was investigated. The Mann-Whitney U test and unpaired t-test were used, in addition to calculating the non-parametric Spearman's correlation coefficient, to statistically evaluate the acquired data. / Results: The hips in this study were found to have had a median unworn diametrical clearance of 90.31 μm (interquartile range (IQR) 77.59 to 97.40); 32% (n = 26) were found to have been below the manufacturing tolerance. There was no correlation found between clearance and bearing (rs = -0.0004, p = 0.997) or taper (rs = 0.0048, p = 0.966) wear rates. The wear performance of hips manufactured within and below these specifications was not significantly different (bearing: p = 0.395; taper: p = 0.653). Pinnacles manufactured from 2007 onwards had a greater prevalence of bearing clearance below tolerance (p = 0.004). / Conclusion: The diametrical clearance of Pinnacle hips did not influence their wear performance, even when below the manufacturing tolerance. The optimum clearance for minimizing hip implant wear remains unclear.

Type: Article
Title: Does diametrical clearance influence the wear of Pinnacle hip implants?
Open access status: An open access version is available from UCL Discovery
DOI: 10.1302/2046-3758.98.BJR-2019-0195.R2
Publisher version: https://doi.org/10.1302/2046-3758.98.BJR-2019-0195...
Language: English
Additional information: Copyright © 2020 Author(s) et al. This is an open-access article distributed under the terms of the Creative Commons Attribution Non-Commercial No Derivatives (CC BY-NC-ND 4.0) licence, which permits the copying and redistribution of the work only, and provided the original author and source are credited. See https://creativecommons.org/licenses/by-nc-nd/4.0/.
UCL classification: UCL
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Medical Sciences
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Medical Sciences > Div of Surgery and Interventional Sci
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Medical Sciences > Div of Surgery and Interventional Sci > Department of Ortho and MSK Science
URI: https://discovery.ucl.ac.uk/id/eprint/10110077
Downloads since deposit
81Downloads
Download activity - last month
Download activity - last 12 months
Downloads by country - last 12 months

Archive Staff Only

View Item View Item