Μulti-institutional dosimetric delivery assessment of intracranial stereotactic radiosurgery on different treatment platforms

Dimitriadis, A; Tsang, Y; Thomas, RAS; Palmer, AL; Eaton, D; Lee, J; Patel, R; ... Clark, CH; + view all (2020) Μulti-institutional dosimetric delivery assessment of intracranial stereotactic radiosurgery on different treatment platforms. Radiotherapy and Oncology , 147 pp. 153-161. 10.1016/j.radonc.2020.05.024. Green open access

[thumbnail of Nisbet_Μulti-institutional dosimetric delivery assessment of intracranial stereotactic radiosurgery on different treatment platforms_AAM.pdf]
Preview
Text
Nisbet_Μulti-institutional dosimetric delivery assessment of intracranial stereotactic radiosurgery on different treatment platforms_AAM.pdf - Accepted Version

Download (2MB) | Preview

Abstract

BACKGROUND AND PURPOSE: Assessment of dosimetric accuracy of radiosurgery on different treatment platforms. MATERIAL AND METHODS: Thirty-three single fraction treatment plans were assessed at thirty centres using an anthropomorphic head phantom with target and brainstem structures. The target being a single irregular shaped target, ∼8cc, 10 mm from the brainstem. The phantom was "immobilised", scanned, planned and treated following the local protocols. EBT-XD films and alanine pellets were used to measure absolute dose, inside both the target and the brainstem, and compared with TPS predicted dose distributions. RESULTS: PTV alanine measurements from gantry-based linacs showed a median percentage difference to the TPS of 0.65%. Cyberknife (CK) had the highest median difference of 2.3% in comparison to the other platforms. GammaKnife (GK) showed the smallest median of 0.3%. Similar trends were observed in the OAR with alanine measurements showing median percentage differences of1.1%, 2.0% and 0.4%, for gantry-based linacs, CK and GK respectively. All platforms showed comparable gamma passing rates between axial and sagittal films. CONCLUSIONS: This comparison has highlighted the dosimetric variation between measured and TPS calculated dose for each delivery platforms.. The results suggest that clinically acceptable agreement with the predicted dose distributions is achievable by all treatment delivery systems. Radiosurgery, Dosimetry, End-to-end, Audit, Anthropomorphic Phantom, Alanine, Radiochromic film.

Type: Article
Title: Μulti-institutional dosimetric delivery assessment of intracranial stereotactic radiosurgery on different treatment platforms
Location: Ireland
Open access status: An open access version is available from UCL Discovery
DOI: 10.1016/j.radonc.2020.05.024
Publisher version: https://doi.org/10.1016/j.radonc.2020.05.024
Language: English
Additional information: This version is the author accepted manuscript. For information on re-use, please refer to the publisher's terms and conditions.
Keywords: Radiosurgery, Dosimetry, Audit, Anthropomorphic phantom, Alanine, Radiochromic film
UCL classification: UCL
UCL > Provost and Vice Provost Offices > UCL BEAMS
UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science
UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science > Dept of Med Phys and Biomedical Eng
URI: https://discovery.ucl.ac.uk/id/eprint/10100155
Downloads since deposit
Loading...
80Downloads
Download activity - last month
Loading...
Download activity - last 12 months
Loading...
Downloads by country - last 12 months
Loading...

Archive Staff Only

View Item View Item