UCL Discovery
UCL home » Library Services » Electronic resources » UCL Discovery

A randomized trial evaluating the accuracy of AF detection by four external ambulatory ECG monitors compared to permanent pacemaker AF detection

Eysenck, W; Freemantle, N; Sulke, N; (2020) A randomized trial evaluating the accuracy of AF detection by four external ambulatory ECG monitors compared to permanent pacemaker AF detection. Journal of Interventional Cardiac Electrophysiology , 57 pp. 361-369. 10.1007/s10840-019-00515-0. Green open access

[thumbnail of Article]
Preview
Text (Article)
Freemantle_MAMMBA Paper.pdf - Accepted Version

Download (490kB) | Preview
[thumbnail of Figure]
Preview
Text (Figure)
Freemantle_Figure 1.pdf - Accepted Version

Download (104kB) | Preview
[thumbnail of Table]
Preview
Text (Table)
Freemantle_Table 1.pdf - Accepted Version

Download (346kB) | Preview

Abstract

Purpose: Several external cardiac monitors (ECMs) have recently been developed. These have never been compared to ‘gold standard’ monitoring with concurrently implanted DDDRP pacemakers. The accuracy of AF detection of Zio XT Monitor (ZM), NUUBO Vest (NV) and Carnation Ambulatory Monitor (CAM) compared with Novacor ‘R’ Test 4 (RT) in patients (pts) with DDDRP PPM advanced Holters as the comparator, was evaluated. Methods: Twenty-one pts. with AF and a DDDRP PPM, each acting as their own control subject, wore every ECM for 2 weeks in randomized order. PPM downloads were performed at application and removal. Device ECGs were compared for AF burden and individual AF episodes with PPM Holters. Pt acceptability, wear time, costs and time expenditure were evaluated. Results: RT AF burden was less accurate than the ZM, NV or CAM (p < 0.05). Probability of inaccurate AF diagnosis was higher for RT than ZM or CAM OR 12.31 and 5.85, respectively (p = 0.025 and p = 0.042). ZM wear time was longer than the RT: 307 h vs. 224 h; p = 0.02. Acceptability was greater for CAM than RT (1.86 ± 2.63 compared with 0.57 ± 1.17 for CAM; p = 0.024). All ECMs were more expensive than RT (p < 0.00001). Conclusions: All new ECMs were more expensive than the RT system; however, the ZM, NV and CAM are all more accurate than current standard practice RT device in AF burden assessment. The RT is more likely to give inaccurate diagnoses than ZM or CAM. This may have clinical implications.

Type: Article
Title: A randomized trial evaluating the accuracy of AF detection by four external ambulatory ECG monitors compared to permanent pacemaker AF detection
Open access status: An open access version is available from UCL Discovery
DOI: 10.1007/s10840-019-00515-0
Publisher version: https://doi.org/10.1007/s10840-019-00515-0
Language: English
Additional information: This version is the author accepted manuscript. For information on re-use, please refer to the publisher’s terms and conditions.
UCL classification: UCL
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Population Health Sciences
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Population Health Sciences > Inst of Clinical Trials and Methodology
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Population Health Sciences > Inst of Clinical Trials and Methodology > Comprehensive CTU at UCL
URI: https://discovery.ucl.ac.uk/id/eprint/10098023
Downloads since deposit
378Downloads
Download activity - last month
Download activity - last 12 months
Downloads by country - last 12 months

Archive Staff Only

View Item View Item