UCL Discovery
UCL home » Library Services » Electronic resources » UCL Discovery

Practical issues and limitations of brain attenuation correction on a simultaneous PET-MR scanner

Mackewn, JE; Stirling, J; Jeljeli, S; Gould, SM; Johnstone, RI; Merida, I; Pike, LC; ... Marsden, PK; + view all (2020) Practical issues and limitations of brain attenuation correction on a simultaneous PET-MR scanner. EJNMMI Physics , 7 (1) , Article 24. 10.1186/s40658-020-00295-x. Green open access

[thumbnail of document.pdf]
Preview
Text
document.pdf - Published Version

Download (1MB) | Preview

Abstract

BACKGROUND: Despite the advent of clinical PET-MR imaging for routine use in 2011 and the development of several methods to address the problem of attenuation correction, some challenges remain. We have identified and investigated several issues that might affect the reliability and accuracy of current attenuation correction methods when these are implemented for clinical and research studies of the brain. These are (1) the accuracy of converting CT Hounsfield units, obtained from an independently acquired CT scan, to 511 keV linear attenuation coefficients; (2) the effect of padding used in the MR head coil; (3) the presence of close-packed hair; (4) the effect of headphones. For each of these, we have examined the effect on reconstructed PET images and evaluated practical mitigating measures. RESULTS: Our major findings were (1) for both Siemens and GE PET-MR systems, CT data from either a Siemens or a GE PET-CT scanner may be used, provided the conversion to 511 keV μ-map is performed by the PET-MR vendor’s own method, as implemented on their PET-CT scanner; (2) the effect of the head coil pads is minimal; (3) the effect of dense hair in the field of view is marked (> 10% error in reconstructed PET images); and (4) using headphones and not including them in the attenuation map causes significant errors in reconstructed PET images, but the risk of scanning without them may be acceptable following sound level measurements. CONCLUSIONS: It is important that the limitations of attenuation correction in PET-MR are considered when designing research and clinical PET-MR protocols in order to enable accurate quantification of brain PET scans. Whilst the effect of pads is not significant, dense hair, the use of headphones and the use of an independently acquired CT-scan can all lead to non-negligible effects on PET quantification. Although seemingly trivial, these effects add complications to setting up protocols for clinical and research PET-MR studies that do not occur with PET-CT. In the absence of more sophisticated PET-MR brain attenuation correction, the effect of all of the issues above can be minimised if the pragmatic approaches presented in this work are followed.

Type: Article
Title: Practical issues and limitations of brain attenuation correction on a simultaneous PET-MR scanner
Open access status: An open access version is available from UCL Discovery
DOI: 10.1186/s40658-020-00295-x
Publisher version: https://doi.org/10.1186/s40658-020-00295-x
Language: English
Additional information: © 2020 BioMed Central Ltd. This article is licensed under a Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/).
Keywords: PET-MR, Brain, Linear attenuation coefficient, PET-CT, Padding, Hair, Headphones
UCL classification: UCL
UCL > Provost and Vice Provost Offices
UCL > Provost and Vice Provost Offices > UCL BEAMS
UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science
UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science > Dept of Med Phys and Biomedical Eng
URI: https://discovery.ucl.ac.uk/id/eprint/10097592
Downloads since deposit
Loading...
80Downloads
Download activity - last month
Loading...
Download activity - last 12 months
Loading...
Downloads by country - last 12 months
Loading...

Archive Staff Only

View Item View Item