A comprehensive approach for correcting voxel-wise b-value errors in diffusion MRI

Lee, Y; Kettinger, AO; Wilm, BJ; Deichmann, R; Weiskopf, N; Lambert, C; Pruessmann, KP; (2019) A comprehensive approach for correcting voxel-wise b-value errors in diffusion MRI. Magnetic Resonance in Medicine 10.1002/mrm.28078. (In press). Green open access

[thumbnail of Lee_et_al-2019-Magnetic_Resonance_in_Medicine.pdf]
Preview
Text
Lee_et_al-2019-Magnetic_Resonance_in_Medicine.pdf - Published Version

Download (2MB) | Preview

Abstract

PURPOSE: In diffusion MRI, the actual b‐value played out on the scanner may deviate from the nominal value due to magnetic field imperfections. A simple image‐based correction method for this problem is presented. METHODS: The apparent diffusion constant (ADC) of a water phantom was measured voxel‐wise along 64 diffusion directions at b = 1000 s/mm2. The true diffusion constant of water was estimated, considering the phantom temperature. A voxel‐wise correction factor, providing an effective b‐value including any magnetic field deviations, was determined for each diffusion direction by relating the measured ADC to the true diffusion constant. To test the method, the measured b‐value map was used to calculate the corrected voxel‐wise ADC for additionally acquired diffusion data sets on the same water phantom and data sets acquired on a small water phantom at three different positions. Diffusion tensor was estimated by applying the measured b‐value map to phantom and in vivo data sets. RESULTS: The b‐value‐corrected ADC maps of the phantom showed the expected spatial uniformity as well as a marked improvement in consistency across diffusion directions. The b‐value correction for the brain data resulted in a 5.8% and 5.5% decrease in mean diffusivity and angular differences of the primary diffusion direction of 2.71° and 0.73° inside gray and white matter, respectively. CONCLUSION: The actual b‐value deviates significantly from its nominal setting, leading to a spatially variable error in the common diffusion outcome measures. The suggested method measures and corrects these artifacts.

Type: Article
Title: A comprehensive approach for correcting voxel-wise b-value errors in diffusion MRI
Open access status: An open access version is available from UCL Discovery
DOI: 10.1002/mrm.28078
Publisher version: https://doi.org/10.1002/mrm.28078
Language: English
Additional information: This is an open access article under the terms of the Creative Commons Attribution License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited. http://creativecommons.org/licenses/by/4.0/
Keywords: b-value, diffusion gradient, diffusion MRI, diffusion tensor, gradient nonlinearity
UCL classification: UCL
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Brain Sciences
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Brain Sciences > UCL Queen Square Institute of Neurology
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Brain Sciences > UCL Queen Square Institute of Neurology > Imaging Neuroscience
URI: https://discovery.ucl.ac.uk/id/eprint/10089991
Downloads since deposit
0Downloads
Download activity - last month
Download activity - last 12 months
Downloads by country - last 12 months

Archive Staff Only

View Item View Item