Feasibility of using a novel non-invasive ambulatory tibial nerve stimulation device for the home-based treatment of overactive bladder symptoms.
Translational Andrology and Urology
, 7
(6)
pp. 912-919.
10.21037/tau.2018.09.12.
Preview |
Text
Panicker_Feasibility of using a novel non-invasive ambulatory tibial nerve stimulation device for the home-based treatment of overactive bladder symptoms_AAM2.pdf - Accepted Version Download (1MB) | Preview |
Abstract
BACKGROUND: To evaluate safety, acceptability and pilot efficacy of transcutaneous low-frequency tibial nerve stimulation (TNS) using a novel device as home-based neuromodulation. METHODS: In this single-centre pilot study, 48 patients with overactive bladder (OAB) (24 with neurogenic and 24 with idiopathic OAB) were randomized to use a self-applicating ambulatory skin-adhering device stimulating transcutaneously the tibial nerve at 1 Hz for 30 minutes, either once daily or once weekly, for 12 weeks. Changes in OAB symptoms and QoL were measured at baseline, weeks 4, 8, and 12 using validated scoring instruments (ICIQ-OAB and ICIQ-LUTSqol), 3-day bladder diary and a Global Response Assessment (GRA) at week 12. RESULTS: Thirty-four patients completed the study (idiopathic n=15, neurogenic n=19). No significant adverse effects were noted. Patients found the device acceptable. Eighteen patients (53%) reported a moderate or marked improvement in symptoms from the GRA. Between baseline and week-12, ICIQ-OAB part A sub-scores improved from mean (SD) 9.3 (2.5) to 7.5 (3.1), and from 9.1 (1.9) to 5.9 (1.7) in the daily and the weekly arms, respectively. ICIQ-LUTSqol part A sub-scores improved from mean (SD) 51 (12.8) to 44.2 (13.1) and 44.9 (9.0) to 35.9 (8.8) in the daily and the weekly arms, respectively. Bladder diary mean 24-hour frequency episodes improved from 11.5 to 8.8 at week 12 for both arms. CONCLUSIONS: This novel ambulatory transcutaneous TNS (TTNS) device is safe and acceptable for use in patients reporting OAB symptoms as a form of home-based neuromodulation. A larger study however is required to confirm clinical efficacy.
Type: | Article |
---|---|
Title: | Feasibility of using a novel non-invasive ambulatory tibial nerve stimulation device for the home-based treatment of overactive bladder symptoms |
Open access status: | An open access version is available from UCL Discovery |
DOI: | 10.21037/tau.2018.09.12 |
Publisher version: | http://doi.org/10.21037/tau.2018.09.12 |
Language: | English |
Additional information: | This version is the author accepted manuscript. For information on re-use, please refer to the publisher’s terms and conditions. |
Keywords: | Overactive bladder (OAB), multiples sclerosis, ranscutaneous tibial nerve stimulation (TTNS) |
UCL classification: | UCL UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Brain Sciences UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Brain Sciences > UCL Queen Square Institute of Neurology |