Zumbo, G;
Barton, SV;
Thompson, D;
Sun, M;
Abdel-Gadir, A;
Treibel, TA;
Knight, D;
... Fontana, M; + view all
(2018)
Extracellular volume with bolus-only technique in amyloidosis patients: Diagnostic accuracy, correlation with other clinical cardiac measures, and ability to track changes in amyloid load over time.
Journal of Magnetic Resonance Imaging
, 47
(6)
pp. 1677-1684.
10.1002/jmri.25907.
Preview |
Text
Abdel-Gadir ECV-with-bolus-only-technique-in-amyloidosis-patients-EIC-edits .docx unblinded.pdf - Accepted Version Download (351kB) | Preview |
Abstract
BACKGROUND: Extracellular volume (ECV) by T1 mapping requires the contrast agent distribution to be at equilibrium. This can be achieved either definitively with a primed contrast infusion (infusion ECV), or sufficiently with a delay postbolus (bolus-only ECV). For large ECV, the bolus-only approach measures higher than the infusion ECV, causing some uncertainty in diseases such as amyloidosis. PURPOSE: To characterize the relationship between the bolus-only and current gold-standard infusion ECV in patients with amyloidosis. STUDY TYPE: Bolus-only and infusion ECV were prospectively measured. POPULATION: In all, 186 subjects with systemic amyloidosis attending our clinic and 23 subjects with systemic amyloidosis who were participating in an open-label, two-part, dose-escalation, phase 1 trial. FIELD STRENGTH: Avanto 1.5T, Siemens Medical Solutions, Erlangen, Germany. ASSESSMENT: Bolus-only and infusion ECV were measured in all subjects using shortened modified Look-Locker inversion recovery (ShMOLLI) T1 mapping sequence. STATISTICAL TESTS: Pearson correlation coefficient (r); Bland-Altman; receiver operating characteristic (ROC) curve analysis. Linear regression model with a fractional polynomial transformation. RESULTS: The difference between the bolus-only and infusion myocardial ECV increased as the average of the two measures increased, with the bolus-ECV measuring higher. For an average ECV of 0.4, the difference was 0.013. The 95% limits of agreement for the two methods, after adjustment for the bias, were ±0.056. However, cardiac diagnostic accuracy was comparable (bolus-only vs. infusion ECV area under the curve [AUC] = 0.839 vs. 0.836), as were correlations with other clinical cardiac measures, and, in the trial patients, the ability to track changes in the liver/spleen with therapy. DATA CONCLUSION: In amyloidosis, with large ECVs, the bolus-only technique reads higher than the infusion technique, but clinical performance by any measure is the same. Given the work-flow advantages, these data suggest that the bolus-only approach might be acceptable for amyloidosis, and might support its use as a surrogate endpoint in future clinical trials. LEVEL OF EVIDENCE: 1 Technical Efficacy: Stage 4 J. Magn. Reson. Imaging 2017.
Archive Staff Only
View Item |