(2017)
Validation of non-invasive central blood pressure devices: Artery society task force (abridged) consensus statement on protocol standardization.
Artery Research
, 20
pp. 35-43.
10.1016/j.artres.2017.11.001.
Preview |
Text
Hughes_Abridged Version_Artery Research.pdf - Accepted Version Download (390kB) | Preview |
Abstract
Brachial cuff blood pressure (BP) is clinically important, but may be an inaccurate substitute for central BP. Many non-invasive devices have been developed that purport to estimate central BP from peripheral artery sites, yet with no standardized guidelines; the accuracy testing of these new devices has not been undertaken in a uniform fashion with comparable protocols. This is an abridged paper describing the recommendations reached by an international task force convened to identify issues that need to be addressed and reach consensus relating to methods for assessing and reporting the accuracy (validation) of central BP devices. The recommendations are endorsed by the Association for Research into Arterial Structure and Physiology (ARTERY) Society, as well as the European Society of Hypertension (ESH) Working Group on Arterial Structure and Function, and the ESH Working Group on Blood Pressure Monitoring and Cardiovascular Variability. Researchers interested in validating central BP monitors should read the full version of the statement.
Type: | Article |
---|---|
Title: | Validation of non-invasive central blood pressure devices: Artery society task force (abridged) consensus statement on protocol standardization |
Open access status: | An open access version is available from UCL Discovery |
DOI: | 10.1016/j.artres.2017.11.001 |
Publisher version: | http://doi.org/10.1016/j.artres.2017.11.001 |
Language: | English |
Additional information: | This version is the author accepted manuscript. For information on re-use, please refer to the publisher’s terms and conditions. |
Keywords: | Guideline, Aorta, Diagnostic equipment, Hypertension, Central blood pressure |
UCL classification: | UCL UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Population Health Sciences > Institute of Cardiovascular Science UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Population Health Sciences > Institute of Cardiovascular Science > Population Science and Experimental Medicine |
URI: | https://discovery.ucl.ac.uk/id/eprint/10037985 |