Collimator design for a clinical brain SPECT/MRI insert

Salvado, D; Erlandsson, K; Bousse, A; Occhipinti, M; Busca, P; Fiorini, C; Hutton, B; (2014) Collimator design for a clinical brain SPECT/MRI insert. EJNMMI Phys , 1 (Suppl) , Article A21. 10.1186/2197-7364-1-S1-A21. Green open access

[thumbnail of Collimator design for a clinical brain SPECT/MRI insert.pdf]
Preview
Text
Collimator design for a clinical brain SPECT/MRI insert.pdf

Download (617kB) | Preview

Abstract

This project's goal is to design a SPECT insert for a clinical MRI system for simultaneous brain SPECT/MR imaging. We assume the stationary SPECT insert will consist of two rings of ∼5x5-cm SiPM-based detectors insensitive to magnetic fields, with 0.8-mm intrinsic resolution. The maximum diameter is 44.5 cm, the minimum diameter is 33 cm to accommodate the patient and MRI receive/transmit coil, and the FOV has a 20 cm diameter. We have compared eight collimator designs: single-, 2x2-, 3x3- and 5+2½- pinhole, and single-, 2-, 3- and 1+2½-slit slit-slat, where ½-pinholes/slits are shared between two detectors. Analytical geometric efficiency was calculated for an activity distribution corresponding to a human brain and a target resolution of 10 mm FWHM at the centre of the FOV. Noise-free data were simulated with and without depth-of-interaction (DOI) information, and reconstructed for uniform, Defrise, Derenzo, and Zubal brain phantoms. For DOI it is assumed that the crystal's first and second half can be differentiated. Comparing the multi-pinhole and multi-slit slit-slat collimators, the former gives better reconstructed uniformity and trans-axial resolution, while the latter gives better axial resolution. Although the 2x2-pinhole and 2-slit designs give the highest sensitivities, they result in a sub-optimal utilization of the detector FOV. The best options are therefore the 5+2½-pinhole and the 1+2½-slit systems, with sensitivities of 4.9*10–4 and 4.0*10–4, respectively. The brain phantom reconstructions with multi-pinhole collimator are superior as compared to slit-slat, especially in terms of symmetry and realistic activity distribution. DOI information reduces artefacts and improves uniformity in geometric phantoms, although the difference is small for the brain phantom. These results favour a multi-pinhole configuration.

Type: Article
Title: Collimator design for a clinical brain SPECT/MRI insert
Location: Germany
Open access status: An open access version is available from UCL Discovery
DOI: 10.1186/2197-7364-1-S1-A21
Publisher version: http://dx.doi.org/10.1186/2197-7364-1-S1-A21
Language: English
Additional information: © 2014 Salvado et al; licensee Springer This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
UCL classification: UCL
UCL > Provost and Vice Provost Offices
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Medical Sciences
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Medical Sciences > Div of Medicine
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Medical Sciences > Div of Medicine > Department of Imaging
URI: https://discovery.ucl.ac.uk/id/eprint/1472955
Downloads since deposit
104Downloads
Download activity - last month
Download activity - last 12 months
Downloads by country - last 12 months

Archive Staff Only

View Item View Item