UCL Discovery
UCL home » Library Services » Electronic resources » UCL Discovery

A simple, realistic walled phantom for intravascular and intracardiac applications

Nisar, H; Moore, J; Piazza, R; Maneas, E; Chen, ECS; Peters, TM; (2020) A simple, realistic walled phantom for intravascular and intracardiac applications. International Journal of Computer Assisted Radiology and Surgery 10.1007/s11548-020-02201-3. (In press). Green open access

[thumbnail of Nisar2020_Article_ASimpleRealisticWalledPhantomF.pdf]
Preview
Text
Nisar2020_Article_ASimpleRealisticWalledPhantomF.pdf - Published Version

Download (2MB) | Preview

Abstract

Purpose: This work aims to develop a simple, anatomically and haptically realistic vascular phantom, compatible with intravascular and intracardiac ultrasound. The low-cost, dual-layered phantom bridges the gap between traditional wall-only and wall-less phantoms by showing both the vessel wall and surrounding tissue in ultrasound imaging. This phantom can better assist clinical tool training, testing of intravascular devices, blood flow studies, and validation of algorithms for intravascular and intracardiac surgical systems. / Methods: Polyvinyl alcohol cryogel (PVA-c) incorporating a scattering agent was used to obtain vessel and tissue-mimicking materials. Our specific design targeted the inferior vena cava and renal bifurcations which were modelled using CAD software. A custom mould and container were 3D-printed for shaping the desired vessel wall. Three phantoms were prepared by varying both the concentrations of scattering agent as well as the number of freeze–thaw cycles to which the phantom layers were subjected during the manufacturing process. Each phantom was evaluated using ultrasound imaging using the Foresight™ ICE probe. Geometrical validation was provided by comparing CAD design to a CT scan of the phantom. / Results: The desired vascular phantom was constructed using 2.5% and 0.05% scattering agent concentration in the vessel and tissue-mimicking layers, respectively. Imaging of the three phantoms showed that increasing the number of freeze–thaw cycles did not significantly enhance the image contrast. Comparison of the US images with their CT equivalents resulted in an average error of 0.9mm for the lumen diameter. / Conclusion: The phantom is anatomically realistic when imaged with intracardiac ultrasound and provides a smooth lumen for the ultrasound probe and catheter to manoeuvre. The vascular phantom enables validation of intravascular and intracardiac image guidance systems. The simple construction technique also provides a workflow for designing complex, multi-layered arterial phantoms.

Type: Article
Title: A simple, realistic walled phantom for intravascular and intracardiac applications
Open access status: An open access version is available from UCL Discovery
DOI: 10.1007/s11548-020-02201-3
Publisher version: https://doi.org/10.1007/s11548-020-02201-3
Language: English
Additional information: This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
Keywords: Vascular phantom, Walled phantom, Polyvinyl alcohol cryogel (PVA-c), Intravascular imaging (IVUS), Intracardiac imaging (ICE)
UCL classification: UCL
UCL > Provost and Vice Provost Offices > UCL BEAMS
UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science
UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science > Dept of Med Phys and Biomedical Eng
URI: https://discovery.ucl.ac.uk/id/eprint/10102784
Downloads since deposit
109Downloads
Download activity - last month
Download activity - last 12 months
Downloads by country - last 12 months

Archive Staff Only

View Item View Item