UCL Discovery
UCL home » Library Services » Electronic resources » UCL Discovery

Evaluation of MRI-derived surrogate signals to model respiratory motion

Tran, EH; Eiben, B; Wetscherek, A; Oelfke, U; Meedt, G; Hawkes, DJ; McClelland, JR; (2020) Evaluation of MRI-derived surrogate signals to model respiratory motion. Biomedical Physics & Engineering Express , 6 (4) , Article 045015. 10.1088/2057-1976/ab944c. Green open access

[thumbnail of Eiben_Tran_2020_Biomed._Phys._Eng._Express_6_045015.pdf]
Preview
Text
Eiben_Tran_2020_Biomed._Phys._Eng._Express_6_045015.pdf - Published Version

Download (2MB) | Preview

Abstract

An MR-Linac can provide motion information of tumour and organs-at-risk before, during, and after beam delivery. However, MR imaging cannot provide real-time high-quality volumetric images which capture breath-to-breath variability of respiratory motion. Surrogate-driven motion models relate the motion of the internal anatomy to surrogate signals, thus can estimate the 3D internal motion from these signals. Internal surrogate signals based on patient anatomy can be extracted from 2D cine-MR images, which can be acquired on an MR-Linac during treatment, to build and drive motion models. In this paper we investigate different MRI-derived surrogate signals, including signals generated by applying principal component analysis to the image intensities, or control point displacements derived from deformable registration of the 2D cine-MR images. We assessed the suitability of the signals to build models that can estimate the motion of the internal anatomy, including sliding motion and breath-to-breath variability. We quantitatively evaluated the models by estimating the 2D motion in sagittal and coronal slices of 8 lung cancer patients, and comparing them to motion measurements obtained from image registration. For sagittal slices, using the first and second principal components on the control point displacements as surrogate signals resulted in the highest model accuracy, with a mean error over patients around 0.80 mm which was lower than the in-plane resolution. For coronal slices, all investigated signals except the skin signal produced mean errors over patients around 1 mm. These results demonstrate that surrogate signals derived from 2D cine-MR images, including those generated by applying principal component analysis to the image intensities or control point displacements, can accurately model the motion of the internal anatomy within a single sagittal or coronal slice. This implies the signals should also be suitable for modelling the 3D respiratory motion of the internal anatomy.

Type: Article
Title: Evaluation of MRI-derived surrogate signals to model respiratory motion
Open access status: An open access version is available from UCL Discovery
DOI: 10.1088/2057-1976/ab944c
Publisher version: https://doi.org/10.1088/2057-1976/ab944c
Language: English
Additional information: Original content from this work may be used under the terms of the Creative Commons Attribution 4.0 licence (http://creativecommons.org/licenses/by/4.0). Any further distribution of this work must maintain attribution to the author(s) and the title of the work, journal citation and DOI.
UCL classification: UCL
UCL > Provost and Vice Provost Offices > UCL BEAMS
UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science
UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science > Dept of Med Phys and Biomedical Eng
URI: https://discovery.ucl.ac.uk/id/eprint/10098404
Downloads since deposit
84Downloads
Download activity - last month
Download activity - last 12 months
Downloads by country - last 12 months

Archive Staff Only

View Item View Item