UCL Discovery
UCL home » Library Services » Electronic resources » UCL Discovery

Catheter manipulation analysis for objective performance and technical skills assessment in transcatheter aortic valve implantation

Mazomenos, EB; Chang, PL; Rippel, RA; Rolls, A; Hawkes, DJ; Bicknell, CD; Desjardins, A; ... Stoyanov, D; + view all (2016) Catheter manipulation analysis for objective performance and technical skills assessment in transcatheter aortic valve implantation. International Journal of Computer Assisted Radiology and Surgery , 11 (6) pp. 1121-1131. 10.1007/s11548-016-1391-6. Green open access

[thumbnail of art%3A10.1007%2Fs11548-016-1391-6.pdf]
Preview
Text
art%3A10.1007%2Fs11548-016-1391-6.pdf

Download (1MB) | Preview

Abstract

PURPOSE: Transcatheter aortic valve implantation (TAVI) demands precise and efficient handling of surgical instruments within the confines of the aortic anatomy. Operational performance and dexterous skills are critical for patient safety, and objective methods are assessed with a number of manipulation features, derived from the kinematic analysis of the catheter/guidewire in fluoroscopy video sequences. METHODS: A silicon phantom model of a type I aortic arch was used for this study. Twelve endovascular surgeons, divided into two experience groups, experts ([Formula: see text]) and novices ([Formula: see text]), performed cannulation of the aorta, representative of valve placement in TAVI. Each participant completed two TAVI experiments, one with conventional catheters and one with the Magellan robotic platform. Video sequences of the fluoroscopic monitor were recorded for procedural processing. A semi-automated tracking software provided the 2D coordinates of the catheter/guidewire tip. In addition, the aorta phantom was segmented in the videos and the shape of the entire catheter was manually annotated in a subset of the available video frames using crowdsourcing. The TAVI procedure was divided into two stages, and various metrics, representative of the catheter's overall navigation as well as its relative movement to the vessel wall, were developed. RESULTS: Experts consistently exhibited lower values of procedure time and dimensionless jerk, and higher average speed and acceleration than novices. Robotic navigation resulted in increased average distance to the vessel wall in both groups, a surrogate measure of safety and reduced risk of embolisation. Discrimination of experience level and types of equipment was achieved with the generated motion features and established clustering algorithms. CONCLUSIONS: Evaluation of surgical skills is possible through the analysis of the catheter/guidewire motion pattern. The use of robotic endovascular platforms seems to enable more precise and controlled catheter navigation.

Type: Article
Title: Catheter manipulation analysis for objective performance and technical skills assessment in transcatheter aortic valve implantation
Open access status: An open access version is available from UCL Discovery
DOI: 10.1007/s11548-016-1391-6
Publisher version: http://dx.doi.org/10.1007/s11548-016-1391-6
Language: English
Additional information: Copyright © The Author(s) 2016. This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
Keywords: Catheter motion analysis, Endovascular robotics, Objective skills assessment
UCL classification: UCL
UCL > Provost and Vice Provost Offices > UCL BEAMS
UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science
UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science > Dept of Computer Science
UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science > Dept of Med Phys and Biomedical Eng
URI: https://discovery.ucl.ac.uk/id/eprint/1482456
Downloads since deposit
38Downloads
Download activity - last month
Download activity - last 12 months
Downloads by country - last 12 months

Archive Staff Only

View Item View Item