A comparison of static and dynamic ∆B0 mapping methods for correction of CEST MRI in the presence of temporal B0 field variations

Poblador Rodriguez, E; Moser, P; Dymerska, B; Robinson, S; Schmitt, B; van der Kouwe, A; Gruber, S; ... Bogner, W; + view all (2019) A comparison of static and dynamic ∆B0 mapping methods for correction of CEST MRI in the presence of temporal B0 field variations. Magnetic Resonance in Medicine 10.1002/mrm.27750. (In press). Green open access

[thumbnail of Rodriguez_et_al-2019-Magnetic_Resonance_in_Medicine.pdf]
Preview
Text
Rodriguez_et_al-2019-Magnetic_Resonance_in_Medicine.pdf - Published Version

Download (1MB) | Preview

Abstract

Purpose: To assess the performance, in the presence of scanner instabilities, of three dynamic correction methods which integrate ∆B 0 mapping into the chemical exchange saturation transfer (CEST) measurement and three established static ∆B 0 -correction approaches. Methods: A homogeneous phantom and five healthy volunteers were scanned with a CEST sequence at 7 T. The in vivo measurements were performed twice: first with unaltered system frequency and again applying frequency shifts during the CEST acquisition. In all cases, retrospective voxel-wise ∆B 0 -correction was performed using one intrinsic and two extrinsic [prescans with dual-echo gradient-echo and water saturation shift referencing (WASSR)] static approaches. These were compared with two intrinsic [using phase data directly generated by single-echo or double-echo GRE (gradient-echo) CEST readout (CEST-GRE-2TE)] and one extrinsic [phase from interleaved dual-echo EPI (echo planar imaging) navigator (NAV-EPI-2TE)] dynamic ∆B 0 -correction approaches [allowing correction of each Z-spectral point before magnetization transfer ratio asymmetry (MTR asym) analysis]. Results: All three dynamic methods successfully mapped the induced drift. The intrinsic approaches were affected by the CEST labeling near water (∆ω < |0.3| ppm). The MTR asym contrast was distorted by the frequency drift in the brain by up to 0.21%/Hz when static ∆B 0 -corrections were applied, whereas the dynamic ∆B 0 corrections reduced this to <0.01%/Hz without the need of external scans. The CEST-GRE-2TE and NAV-EPI-2TE resulted in highly consistent MTR asym values with/without drift for all subjects. Conclusion: Reliable correction of scanner instabilities is essential to establish clinical CEST MRI. The three dynamic approaches presented improved the ∆B 0 -correction performance significantly in the presence of frequency drift compared to established static methods. Among them, the self-corrected CEST-GRE-2TE was the most accurate and straightforward to implement.

Type: Article
Title: A comparison of static and dynamic ∆B0 mapping methods for correction of CEST MRI in the presence of temporal B0 field variations
Open access status: An open access version is available from UCL Discovery
DOI: 10.1002/mrm.27750
Publisher version: https://doi.org/10.1002/mrm.27750
Language: English
Additional information: This work is licensed under a Creative Commons Attribution 4.0 International License. The images or other third party material in this article are included in the article’s Creative Commons license, unless indicated otherwise in the credit line; if the material is not included under the Creative Commons license, users will need to obtain permission from the license holder to reproduce the material. To view a copy of this license, visit http://creativecommons.org/licenses/by/4.0/
Keywords: ∆B0 mapping, chemical exchange saturation transfer (CEST), dynamic ∆B0 correction, frequency drift, scanner instabilities
UCL classification: UCL
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Brain Sciences
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Brain Sciences > UCL Queen Square Institute of Neurology
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Brain Sciences > UCL Queen Square Institute of Neurology > Imaging Neuroscience
URI: https://discovery.ucl.ac.uk/id/eprint/10072333
Downloads since deposit
220Downloads
Download activity - last month
Download activity - last 12 months
Downloads by country - last 12 months

Archive Staff Only

View Item View Item