UCL Discovery
UCL home » Library Services » Electronic resources » UCL Discovery

Implant and Midterm Outcomes of the Subcutaneous Implantable Cardioverter-Defibrillator Registry: The EFFORTLESS Study.

Boersma, L; Barr, C; Knops, R; Theuns, D; Eckardt, L; Neuzil, P; Scholten, M; ... EFFORTLESS Investigator Group; + view all (2017) Implant and Midterm Outcomes of the Subcutaneous Implantable Cardioverter-Defibrillator Registry: The EFFORTLESS Study. J Am Coll Cardiol , 70 (7) pp. 830-841. 10.1016/j.jacc.2017.06.040. Green open access

[thumbnail of Lambiase_1-s2.0-S0735109717379287-main.pdf]
Preview
Text
Lambiase_1-s2.0-S0735109717379287-main.pdf - Published Version

Download (1MB) | Preview

Abstract

BACKGROUND: The subcutaneous implantable cardioverter-defibrillator (S-ICD) was developed to defibrillate ventricular arrhythmias, avoiding drawbacks of transvenous leads. The global EFFORTLESS S-ICD (Evaluation oF FactORs ImpacTing CLinical Outcome and Cost EffectiveneSS of the S-ICD) registry is collecting outcomes in 985 patients during a 5-year follow-up. OBJECTIVES: The primary goal of the EFFORTLESS registry is to determine the safety of the S-ICD by evaluating complications and inappropriate shock rate. METHODS: This is the first report on the full patient cohort and study endpoints with follow-up ≥1 year. The predefined endpoints are 30- and 360-day complications, and shocks for atrial fibrillation or supraventricular tachycardia. RESULTS: Patients were followed for 3.1 ± 1.5 years and 82 completed the study protocol 5-year visit. Average age was 48 years, 28% were women, ejection fraction was 43 ± 18%, and 65% had a primary prevention indication. The S-ICD system and procedure complication rate was 4.1% at 30 days and 8.4% at 360 days. The 1-year complication rate trended toward improvement from the first to last quartile of enrollment (11.3% [quartile 1]) to 7.8% [quartile 2], 6.6% [quartile 3], and 7.4% [quartile 4]; quartile 1 vs. quartiles 2 to 4; p = 0.06). Few device extractions occurred due to need for antitachycardia (n = 5), or biventricular (n = 4) or bradycardia pacing (n = 1). Inappropriate shocks occurred in 8.1% at 1 year and 11.7% after 3.1 years. At implant, 99.5% of patients had a successful conversion of induced ventricular tachycardia or ventricular fibrillation. The 1- and 5-year rates of appropriate shock were 5.8% and 13.5%, respectively. Conversion success for discrete spontaneous episodes was 97.4% overall. CONCLUSIONS: This registry demonstrates that the S-ICD fulfills predefined endpoints for safety and efficacy. Midterm performance rates on complications, inappropriate shocks, and conversion efficacy were comparable to rates observed in transvenous implantable cardioverter-defibrillator studies. (Evaluation oF Factors ImpacTing CLinical Outcome and Cost EffectiveneSS of the S-ICD [The EFFORTLESS S-ICD Registry]; NCT01085435).

Type: Article
Title: Implant and Midterm Outcomes of the Subcutaneous Implantable Cardioverter-Defibrillator Registry: The EFFORTLESS Study.
Location: United States
Open access status: An open access version is available from UCL Discovery
DOI: 10.1016/j.jacc.2017.06.040
Publisher version: https://doi.org/10.1016/j.jacc.2017.06.040
Language: English
Additional information: © 2017 The Authors. Published by Elsevier on behalf of the American College of Cardiology Foundation. This is an Open Access article under the CC BY-NC-ND license ( http://creativecommons.org/licenses/by-nc-nd/4.0/ ) .
Keywords: Appropriate therapy, inappropriate therapy, leadless, primary prevention, secondary prevention, ventricular arrhythmias
UCL classification: UCL
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Population Health Sciences > Institute of Cardiovascular Science
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Population Health Sciences > Institute of Cardiovascular Science > Clinical Science
URI: https://discovery.ucl.ac.uk/id/eprint/1571709
Downloads since deposit
38Downloads
Download activity - last month
Download activity - last 12 months
Downloads by country - last 12 months

Archive Staff Only

View Item View Item