Evaluation of the Sysmex XN-550, a Novel Compact Haematology analyser from the XN-L (R) series, compared to the XN-20 system

Tailor, H; Mackie, I; Mellick, A; Machin, S; (2017) Evaluation of the Sysmex XN-550, a Novel Compact Haematology analyser from the XN-L (R) series, compared to the XN-20 system. International Journal of Laboratory Hematology , 39 (6) pp. 585-589. 10.1111/ijlh.12701. Green open access

[thumbnail of Tailor_Evaluation_Sysmex_XN-550.pdf]
Preview
Text
Tailor_Evaluation_Sysmex_XN-550.pdf - Accepted Version

Download (461kB) | Preview

Abstract

INTRODUCTION: The XN-550 is a new, automated, compact, haematology analyser designed to generate a full blood count with a standard five-part white blood cell differential and an immature granulocyte count, as well as an optional reticulocyte and optical platelet (PLT) counts. The aim of the study was to evaluate the performance of the XN-550 and compare it to the established XN-20 system. METHODS: We evaluated the basic parameter and special measurement channels of the XN-550, using the XN-20 (which has a similar operating system), as a reference analyser. Precision, carry-over and throughput evaluations were performed. In addition, a total of 202 samples including normal controls and various pathological samples were studied for comparability. RESULTS: Good correlations with the reference analyser were obtained for all parameters except basophils. The XN-550 offers impedance and optical PLT counts and the latter showed a better correlation and less scatter than the impedance count and was comparable to the XN-20 fluorescent count at PLT counts ≤40×109/L. Precision was good, and no significant carry-over was detected. CONCLUSIONS: The XN-550 was simple and easy to use, while maintaining the good diagnostic sensitivity seen with high-range systems such as the XN-20, making this compact device suitable for near-patient services and smaller satellite laboratories.

Type: Article
Title: Evaluation of the Sysmex XN-550, a Novel Compact Haematology analyser from the XN-L (R) series, compared to the XN-20 system
Open access status: An open access version is available from UCL Discovery
DOI: 10.1111/ijlh.12701
Publisher version: http://dx.doi.org/10.1111/ijlh.12701
Language: English
Additional information: This version is the author accepted manuscript. For information on re-use, please refer to the publisher’s terms and conditions.
Keywords: general hematology; laboratory automation; leukocytes; platelets; reticulocytes; Sysmex
UCL classification: UCL
UCL > Provost and Vice Provost Offices
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Medical Sciences
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Medical Sciences > Cancer Institute
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Medical Sciences > Cancer Institute > Research Department of Haematology
URI: https://discovery.ucl.ac.uk/id/eprint/1562280
Downloads since deposit
3,162Downloads
Download activity - last month
Download activity - last 12 months
Downloads by country - last 12 months

Archive Staff Only

View Item View Item