Dempsey, LA;
Cooper, RJ;
Roque, T;
Correia, T;
Magee, E;
Powell, S;
Gibson, AP;
(2015)
Data-driven approach to optimum wavelength selection for diffuse optical imaging.
Journal of Biomedical Optics
, 20
(1)
, Article 016003. 10.1117/1.JBO.20.1.016003.
Preview |
Text
Hebden_Dempsey_JBO_2015.pdf - Published Version Download (6MB) | Preview |
Abstract
The production of accurate and independent images of the changes in concentration of oxyhemoglobin and deoxyhemoglobin by diffuse optical imaging is heavily dependent on which wavelengths of near-infrared light are chosen to interrogate the target tissue. Although wavelengths can be selected by theoretical methods, in practice the accuracy of reconstructed images will be affected by wavelength-specific and system-specific factors such as laser source power and detector sensitivity. We describe the application of a data-driven approach to optimum wavelength selection for the second generation of University College London's multichannel, time-domain optical tomography system (MONSTIR II). By performing a functional activation experiment using 12 different wavelengths between 690 and 870 nm, we were able to identify the combinations of 2, 3, and 4 wavelengths which most accurately reproduced the results obtained using all 12 wavelengths via an imaging approach. Our results show that the set of 2, 3, and 4 wavelengths which produce the most accurate images of functional activation are [770, 810], [770, 790, 850], and [730, 770, 810, 850] respectively, but also that the system is relatively robust to wavelength selection within certain limits. Although these results are specific to MONSTIR II, the approach we developed can be applied to other multispectral near-infrared spectroscopy and optical imaging systems.
Type: | Article |
---|---|
Title: | Data-driven approach to optimum wavelength selection for diffuse optical imaging |
Open access status: | An open access version is available from UCL Discovery |
DOI: | 10.1117/1.JBO.20.1.016003 |
Publisher version: | http://dx.doi.org/10.1117/1.JBO.20.1.016003 |
Language: | English |
Additional information: | Copyright © 2015 Society of Photo Optical Instrumentation Engineers. One print or electronic copy may be made for personal use only. Systematic reproduction and distribution, duplication of any material in this paper for a fee or for commercial purposes, or modification of the content of the paper are prohibited. |
Keywords: | Diffuse optical tomography, diffuse optical imaging, time resolved, functional near-infrared spectroscopy, wavelength, selection |
UCL classification: | UCL UCL > Provost and Vice Provost Offices > UCL BEAMS UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science > Dept of Med Phys and Biomedical Eng |
URI: | https://discovery.ucl.ac.uk/id/eprint/1545521 |
Archive Staff Only