Yerworth, RJ; Frerichs, I; Bayford, R; (2017) Analysis and compensation for errors in electrical impedance tomography images and ventilation-related measures due to serial data collection. Journal of Clinical Monitoring and Computing , 31 pp. 1093-1101. 10.1007/s10877-016-9920-y. Green open access

[thumbnail of art%3A10.1007%2Fs10877-016-9920-y.pdf]
Preview
Text
art%3A10.1007%2Fs10877-016-9920-y.pdf

Download (1MB) | Preview

Abstract

Electrical impedance tomography (EIT) is increasingly being used as a bedside tool for monitoring regional lung ventilation. However, most clinical systems use serial data collection which, if uncorrected, results in image distortion, particularly at high breathing rates. The objective of this study was to determine the extent to which this affects derived parameters. Raw EIT data were acquired with the GOE-MF II EIT device (CareFusion, Höchberg, Germany) at a scan rate of 13 images/s during both spontaneous breathing and mechanical ventilation. Boundary data for periods of undisturbed tidal breathing were corrected for serial data collection errors using a Fourier based algorithm. Images were reconstructed for both the corrected and original data using the GREIT algorithm, and parameters describing the filling characteristics of the right and left lung derived on a breath by breath basis. Values from the original and corrected data were compared using paired t-tests. Of the 33 data sets, 23 showed significant differences in filling index for at least one region, 11 had significant differences in calculated tidal impedance change and 12 had significantly different filling fractions (p = 0.05). We conclude that serial collection errors should be corrected before image reconstruction to avoid clinically misleading results.

Type: Article
Title: Analysis and compensation for errors in electrical impedance tomography images and ventilation-related measures due to serial data collection
Location: Netherlands
Open access status: An open access version is available from UCL Discovery
DOI: 10.1007/s10877-016-9920-y
Publisher version: http://dx.doi.org/10.1007/s10877-016-9920-y
Language: English
Additional information: © The Author(s) 2016 Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
Keywords: EIT, Imaging, Lung, Mechanical ventilation, Ventilation distribution
UCL classification: UCL
UCL > Provost and Vice Provost Offices
UCL > Provost and Vice Provost Offices > UCL BEAMS
UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science
UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science > Dept of Med Phys and Biomedical Eng
URI: https://discovery.ucl.ac.uk/id/eprint/1508575
Downloads since deposit
Loading...
52Downloads
Download activity - last month
Loading...
Download activity - last 12 months
Loading...
Downloads by country - last 12 months
No data found.

Archive Staff Only

View Item View Item