The 'Sphere': A Dedicated Bifurcation Aneurysm Flow-Diverter Device.

Peach, T; Cornhill, JF; Nguyen, A; Riina, H; Ventikos, Y; (2014) The 'Sphere': A Dedicated Bifurcation Aneurysm Flow-Diverter Device. Cardiovasc Eng Technol , 5 (4) 334 - 347. 10.1007/s13239-014-0188-4. Green open access

[thumbnail of The 'Sphere': A Dedicated Bifurcation Aneurysm Flow-Diverter Device..pdf]
Preview
PDF
The 'Sphere': A Dedicated Bifurcation Aneurysm Flow-Diverter Device..pdf
Available under License : See the attached licence file.

Download (4MB)

Abstract

We present flow-based results from the early stage design cycle, based on computational modeling, of a prototype flow-diverter device, known as the 'Sphere', intended to treat bifurcation aneurysms of the cerebral vasculature. The device is available in a range of diameters and geometries and is constructed from a single loop of NITINOL(®) wire. The 'Sphere' reduces aneurysm inflow by means of a high-density, patterned, elliptical surface that partially occludes the aneurysm neck. The device is secured in the healthy parent vessel by two armatures in the shape of open loops, resulting in negligible disruption of parent or daughter vessel flow. The device is virtually deployed in six anatomically accurate bifurcation aneurysms: three located at the Basilar tip and three located at the terminus bifurcation of the Internal Carotid artery (at the meeting of the middle cerebral and anterior cerebral arteries). Both steady state and transient flow simulations reveal that the device presents with a range of aneurysm inflow reductions, with mean flow reductions falling in the range of 30.6-71.8% across the different geometries. A significant difference is noted between steady state and transient simulations in one geometry, where a zone of flow recirculation is not captured in the steady state simulation. Across all six aneurysms, the device reduces the WSS magnitude within the aneurysm sac, resulting in a hemodynamic environment closer to that of a healthy vessel. We conclude from extensive CFD analysis that the 'Sphere' device offers very significant levels of flow reduction in a number of anatomically accurate aneurysm sizes and locations, with many advantages compared to current clinical cylindrical flow-diverter designs. Analysis of the device's mechanical properties and deployability will follow in future publications.

Type: Article
Title: The 'Sphere': A Dedicated Bifurcation Aneurysm Flow-Diverter Device.
Open access status: An open access version is available from UCL Discovery
DOI: 10.1007/s13239-014-0188-4
Publisher version: http://dx.doi.org/10.1007/s13239-014-0188-4
Additional information: Copyright © 2014 The Author(s). This article is published with open access at Springerlink.com
Keywords: CFD, Medical devices, Minimally invasive, Neurovascular, PED, SILK, Shear stress, Stent, Thrombosis, WEB, WSS
UCL classification: UCL
UCL > Provost and Vice Provost Offices > UCL BEAMS
UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science
UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science > Dept of Mechanical Engineering
URI: https://discovery.ucl.ac.uk/id/eprint/1465428
Downloads since deposit
120Downloads
Download activity - last month
Download activity - last 12 months
Downloads by country - last 12 months

Archive Staff Only

View Item View Item