The automation of routine light transmission platelet aggregation

Lawrie, AS; Lane, PJ; Mackie, IJ; Machin, SJ; Kobayashi, K; (2014) The automation of routine light transmission platelet aggregation. International Journal of Laboratory Hematology , 36 (4) 431 - 438. 10.1111/ijlh.12161. Green open access

[thumbnail of cen12265.pdf]
Preview
PDF
cen12265.pdf

Download (200kB)

Abstract

Introduction: The investigation of platelet function by aggregometry requires specialist equipment and is labour intensive. We have developed an automated platelet aggregation method on a routine coagulation analyser. Methods: We used a CS-2000i (Sysmex) with prototype software to perform aggregation in platelet-rich plasma (PRP), using the following agonists: ADP (0.5-10 μm), epinephrine (0.5-10 μm), collagen (0.5-10 mg/μL), ristocetin (0.75-1.25 mg/mL) and arachidonic acid (0.12-1.0 mm). Platelet agonists were from Hyphen Biomed, and an AggRAM aggregometer (Helena Biosciences) was used as the reference instrument. Results: CS-2000i reaction cuvette stirrer speed was found to influence reaction sensitivity and was optimized to 800 rpm. There were no clinically significant changes in aggregation response when the PRP platelet count was 150-480 x 10/L, but below this there were changes in the maximum amplitude (MA) and slope (rate). Dose response with each of the agonists was comparable between CS-2000i and an AggRAM aggregometer and normal subjects receiving antiplatelet drugs. Aggregation imprecision was similar on both the CS-2000i and AggRAM systems, with a cv for 2-5 μm ADP MA and slope varying between 3-12%. Conclusion: Our preliminary studies indicated that optimal sensitivity using the CS-2000i was obtained with a reaction cuvette stirrer speed of 800 rpm and a PRP platelet count of 200-300 x 10/L; aggregation with a PRP count <100 x 10/L showed poor sensitivity. Imprecision and detection of antiplatelet drug effects was similar between the CS-2000i and AggRAM. These data demonstrate that CS-2000i is comparable to a stand-alone aggregometer, although CS-2000i has the advantages of walk-away technology and also required a smaller sample volume than the AggRAM (44% less). © 2013 John Wiley & Sons Ltd.

Type: Article
Title: The automation of routine light transmission platelet aggregation
Open access status: An open access version is available from UCL Discovery
DOI: 10.1111/ijlh.12161
Publisher version: http://dx.doi.org/10.1111/ijlh.12161
Additional information: © 2013 The Authors. International Journal of Laboratory Hematology Published by John Wiley & Sons Ltd. This is an open access article under the terms of the Creative Commons Attribution License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
Keywords: Automation; platelet aggregation;
UCL classification: UCL
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Medical Sciences
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Medical Sciences > Cancer Institute
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Medical Sciences > Cancer Institute > Research Department of Haematology
URI: https://discovery.ucl.ac.uk/id/eprint/1436561
Downloads since deposit
167Downloads
Download activity - last month
Download activity - last 12 months
Downloads by country - last 12 months

Archive Staff Only

View Item View Item