Yao, Junke;
Pi, Xinyi;
Bosi, Giorgia Maria;
Burriesci, Gaetano;
Wurdemann, Helge;
(2025)
Synchronous Inflation of a Valvuloplasty Balloon Catheter with Heart
Rate: In-vitro Evaluation in Terms of Dilatation Performance.
IEEE Robotics and Automation Letters
(In press).
Preview |
Text
RAL2023_BalloonSychronization_JunkeYao_60bpmOnly_revised.pdf - Accepted Version Download (9MB) | Preview |
Abstract
Balloon aortic valvuloplasty (BAV), a minimally invasive procedure to alleviate aortic valve stenosis, commonly employs rapid ventricular pacing (RVP) for balloon stabilization. However, the repeated and extended operation time associated with this technique poses potential complications. This paper introduces a novel approach to mitigate these concerns by employing a dilatation mechanism that is synchronized with the cardiac frequency, wherein the balloon catheter is fully inflated and deflated to a safe, low volume during the decrement of the ventricular pressure. The synchronized pacing was tested at a heart rate of 60 bpm. To experimentally validate the performance of this new approach, mock aortic roots reproducing different calcification patterns were used to compare the leaflets’ mobility after the dilatation test with traditional BAV. Results confirm successful balloon pacing, maintaining low volume before the ventricular pressure increases. The dilatation performance assessment underscores that the proposed methodology resulted in a higher improvement in terms of the transvalvular pressure gradient and opening area. Optimal performance occurs at 60 bpm, yielding a 30.28% gradient decrease and a 21.35% opening area increase. This research represents a notable step forward toward the development of BAV devices capable of autonomous stabilization, eliminating the need for RVP and its related complications. Furthermore, the use of calcified aortic root (AR) phantoms contributes to an enhanced understanding of hemodynamic implications during BAV procedures.
Type: | Article |
---|---|
Title: | Synchronous Inflation of a Valvuloplasty Balloon Catheter with Heart Rate: In-vitro Evaluation in Terms of Dilatation Performance |
Open access status: | An open access version is available from UCL Discovery |
Publisher version: | https://ieeexplore.ieee.org/xpl/RecentIssue.jsp?pu... |
Language: | English |
Additional information: | This version is the author accepted manuscript. For information on re-use, please refer to the publisher’s terms and conditions. |
UCL classification: | UCL UCL > Provost and Vice Provost Offices > UCL BEAMS UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science > Dept of Mechanical Engineering |
URI: | https://discovery.ucl.ac.uk/id/eprint/10200817 |
Archive Staff Only
![]() |
View Item |