In-Depth Retinal Sensitivity Assessment With the MP3 Type S Microperimeter: A Methods Study

de Guimaraes, Thales AC; de Guimaraes, Isabela MC; Ali, Naser; Kalitzeos, Angelos; Michaelides, Michel; (2024) In-Depth Retinal Sensitivity Assessment With the MP3 Type S Microperimeter: A Methods Study. TVST , 13 (4) , Article 14. 10.1167/tvst.13.4.14. Green open access

[thumbnail of Kalitzeos_In-Depth Retinal Sensitivity Assessment With the MP3 Type S Microperimeter_VoR.pdf]
Preview
PDF
Kalitzeos_In-Depth Retinal Sensitivity Assessment With the MP3 Type S Microperimeter_VoR.pdf - Published Version

Download (4MB) | Preview

Abstract

PURPOSE: Retinal sensitivity is frequently listed as an end point in clinical trials, often with long working practices. The purpose of this methods study was to provide a new workflow and reduced test time for in-depth characterization of retinal sensitivity. METHODS: A workflow for the MP3-S microperimeter with detailed functional characterization of the retina under photopic, mesopic, and scotopic conditions was evaluated. Grids of 32 and 28 test positions for photopic/mesopic and scotopic, respectively, were tested in 12 healthy individuals and compared with an established 68-point grid for test time, mean sensitivity (MS), and bivariate contour ellipse area (BCEA). RESULTS: The mean test time (range; ±SD) was 10.5 minutes (8.4-14.9; ±2.0) in the 68-point grid and 4.3 minutes (3.8-5.0; ±0.4) in the 32-point grid, which was significantly different (P < 0.0001). The mean of difference in test time (±SD; 95% confidence interval) was 6.1 minutes (±2.0; 4.6-7.6). MS and BCEA were significantly correlated between grids (r = 0.89 and 0.74; P = 0.0005 and 0.014, respectively). Mean test time of subjects who underwent the full protocol (n = 4) was 2.15 hours. CONCLUSIONS: The protocol suggested herein appears highly feasible with in-depth characterization of retinal function under different testing conditions and in a short test time. TRANSLATIONAL RELEVANCE: The protocol described herein allows for characterization of the retina under different testing conditions and in a short test time, which is relevant due to its potential for patient prognostication and follow-up in clinical settings and also given its increasing role as a clinical trial end point.

Type: Article
Title: In-Depth Retinal Sensitivity Assessment With the MP3 Type S Microperimeter: A Methods Study
Location: United States
Open access status: An open access version is available from UCL Discovery
DOI: 10.1167/tvst.13.4.14
Publisher version: http://dx.doi.org/10.1167/tvst.13.4.14
Language: English
Additional information: This work is licensed under a Creative Commons Attribution 4.0 International License.
Keywords: Humans, Retina
UCL classification: UCL
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Brain Sciences
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Brain Sciences > Institute of Ophthalmology
URI: https://discovery.ucl.ac.uk/id/eprint/10190849
Downloads since deposit
6Downloads
Download activity - last month
Download activity - last 12 months
Downloads by country - last 12 months

Archive Staff Only

View Item View Item