Jaccottet, Arthur;
Feng, Peilong;
Szostak-Lipowicz, Katarzyna M;
Keeble, Lewis;
Constandinou, Timothy G;
(2022)
Towards a wireless micropackaged implant with hermeticity monitoring.
In:
Proceedings of the IEEE Biomedical Circuits and Systems Conference (BioCAS) 2022.
(pp. pp. 500-504).
Institute of Electrical and Electronics Engineers (IEEE)
Preview |
Text
Jaccottet_Towards a wireless micropackaged implant with hermeticity monitoring_AAM.pdf - Accepted Version Download (1MB) | Preview |
Abstract
The development of reliable hermetic chip-scale micropackaging is one of the major challenges in the miniaturization of implantable medical devices. Protecting the patient from the implanted foreign body and the implant itself from the biological environment is crucial. This paper presents an implantable micropackaging concept to protect a microelectronic system-on-chip. A hermetic chamber is formed by bonding the active CMOS chip to a silicon cover using a gold-tin eutectic sealant. The cover’s fabrication method and the die’s post-processing steps are presented. A humidity sensor inside the chamber monitors the humidity to assess permeability. To power the sensor and read its data, interconnections in the CMOS chip have been designed; these metal tracks pass underneath the cover and thus create a connection between the inside and the outside of the cavity. As an alternative to these connections, an on-chip wireless power management and data communication system is presented with simulated results.
Type: | Proceedings paper |
---|---|
Title: | Towards a wireless micropackaged implant with hermeticity monitoring |
Event: | 2022 IEEE Biomedical Circuits and Systems Conference (BioCAS) |
Location: | Taipei, Taiwan |
Dates: | 13th-15th October 2022 |
ISBN-13: | 978-1-6654-6917-3 |
Open access status: | An open access version is available from UCL Discovery |
DOI: | 10.1109/biocas54905.2022.9948556 |
Publisher version: | http://dx.doi.org/10.1109/biocas54905.2022.9948556 |
Language: | English |
Additional information: | This version is the author accepted manuscript. For information on re-use, please refer to the publisher's terms and conditions. |
UCL classification: | UCL UCL > Provost and Vice Provost Offices > UCL BEAMS UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science > Dept of Mechanical Engineering |
URI: | https://discovery.ucl.ac.uk/id/eprint/10184429 |
Archive Staff Only