UCL Discovery
UCL home » Library Services » Electronic resources » UCL Discovery

Design of an Ultra-low-dose, Stationary, Tomographic Molecular Breast Imaging System

Hutton, BF; Erlandsson, K; Wirth, A; Baistow, I; Thielemans, K; Cherlin, A; (2022) Design of an Ultra-low-dose, Stationary, Tomographic Molecular Breast Imaging System. In: 2021 IEEE Nuclear Science Symposium and Medical Imaging Conference Record, NSS/MIC 2021 and 28th International Symposium on Room-Temperature Semiconductor Detectors, RTSD 2022. IEEE: Piscataway, NJ, USA. Green open access

[thumbnail of MIC21_proc_MBI_BH_noc.pdf]
Preview
Text
MIC21_proc_MBI_BH_noc.pdf - Accepted Version

Download (353kB) | Preview

Abstract

Molecular Breast Imaging (MBI) has been shown to have high sensitivity in detection of cancer, even in patients with dense breasts where conventional mammography has issues. However the technique has limited acceptance due to the relatively high radiation dose and long imaging time. Improved lesion detection can be achieved using tomography, however this normally involves detector motion and complex mechanics. Our goal is to develop a low-dose stationary tomographic MBI system with similar or better sensitivity for lesion detection to conventional planar MBI. The proposed system utilizes state-of-the-art cadmium zinc telluride (CZT) detectors based on 2mm pixels, with sub-pixelization and depth of interaction (DOI) capability, combined with densely packed multi-pinhole collimators. Use of closely-spaced pinholes improves efficiency and angular sampling, but results in significant multiplexing. De-multiplexing algorithms have been developed that take advantage of the DOI acquisition to achieve tomographic reconstruction using two opposing planar detectors which apply mild compression to the breast. Simulation studies of multiple lesions with clinically realistic contrast have been used to demonstrate the feasibility of the design and to characterize the expected performance. Reconstruction without de-multiplexing resulted in significant artefacts. De-multiplexing without DOI had limited success but with DOI resulted in artefact-free images, with good contrast and axial plane definition. Lesion detectability was preserved even with reduction of acquisition time (or radiation dose) by a factor of 4. Further optimization has potential for even greater dose reduction. A prototype system is currently being constructed to validate these findings.

Type: Proceedings paper
Title: Design of an Ultra-low-dose, Stationary, Tomographic Molecular Breast Imaging System
Event: 2021 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC)
Dates: 16 Oct 2021 - 23 Oct 2021
Open access status: An open access version is available from UCL Discovery
DOI: 10.1109/NSS/MIC44867.2021.9875933
Publisher version: https://doi.org/10.1109/NSS/MIC44867.2021.9875933
Language: English
Additional information: This version is the author accepted manuscript. For information on re-use, please refer to the publisher’s terms and conditions.
Keywords: Multiplexing , Sensitivity , Prototypes , Detectors , Tomography , Lesions , Image reconstruction
UCL classification: UCL
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Medical Sciences
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Medical Sciences > Div of Medicine
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Medical Sciences > Div of Medicine > Department of Imaging
URI: https://discovery.ucl.ac.uk/id/eprint/10159278
Downloads since deposit
60Downloads
Download activity - last month
Download activity - last 12 months
Downloads by country - last 12 months

Archive Staff Only

View Item View Item