Robson, Anthony G;
Frishman, Laura J;
Grigg, John;
Hamilton, Ruth;
Jeffrey, Brett G;
Kondo, Mineo;
Li, Shiying;
(2022)
ISCEV Standard for full-field clinical electroretinography (2022 update).
Documenta Ophthalmologica
, 144
(3)
pp. 165-177.
10.1007/s10633-022-09872-0.
Preview |
PDF
ISCEV Standard for full-field clinical electroretinography (2022 update).pdf - Other Download (598kB) | Preview |
Abstract
The full-field electroretinogram (ERG) is a mass electrophysiological response to diffuse flashes of light and is used widely to assess generalized retinal function. This document, from the International Society for Clinical Electrophysiology of Vision (ISCEV), presents an updated and revised ISCEV Standard for clinical ERG testing. Minimum protocols for basic ERG stimuli, recording methods and reporting are specified, to promote consistency of methods for diagnosis, monitoring and inter-laboratory comparisons, while also responding to evolving clinical practices and technology. The main changes in this updated ISCEV Standard for clinical ERGs include specifying that ERGs may meet the Standard without mydriasis, providing stimuli adequately compensate for non-dilated pupils. There is more detail about analysis of dark-adapted oscillatory potentials (OPs) and the document format has been updated and supplementary content reduced. There is a more detailed review of the origins of the major ERG components. Several tests previously tabulated as additional ERG protocols are now cited as published ISCEV extended protocols. A non-standard abbreviated ERG protocol is described, for use when patient age, compliance or other circumstances preclude ISCEV Standard ERG testing.
Type: | Article |
---|---|
Title: | ISCEV Standard for full-field clinical electroretinography (2022 update) |
Location: | Netherlands |
Open access status: | An open access version is available from UCL Discovery |
DOI: | 10.1007/s10633-022-09872-0 |
Publisher version: | https://doi.org/10.1007/s10633-022-09872-0 |
Language: | English |
Additional information: | This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. |
Keywords: | Clinical standards, Electroretinogram (ERG), Full-feld ERG, International Society of Clinical Electrophysiology of Vision (ISCEV) |
UCL classification: | UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Brain Sciences UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Brain Sciences > Institute of Ophthalmology UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences UCL |
URI: | https://discovery.ucl.ac.uk/id/eprint/10154215 |
Archive Staff Only
View Item |