UCL Discovery
UCL home » Library Services » Electronic resources » UCL Discovery

Comparison of noise reduction methods in photoacoustic microscopy

Guney, G; Uluc, N; Demirkiran, A; Aytac-Kipergil, E; Unlu, MB; Birgul, O; (2019) Comparison of noise reduction methods in photoacoustic microscopy. Computers in Biology and Medicine , 109 pp. 333-341. 10.1016/j.compbiomed.2019.04.035. Green open access

[thumbnail of Aytac Kipergil_CBM-D-19-00022_R2-26-57.pdf]
Preview
Text
Aytac Kipergil_CBM-D-19-00022_R2-26-57.pdf - Accepted Version

Download (1MB) | Preview

Abstract

Photoacoustic microscopy (PAM) is classified as a hybrid imaging technique based on the photoacoustic effect and has been frequently studied in recent years. Photoacoustic (PA) signals are inherently recorded in a noisy environment and are also exposed to noise by system components. Therefore, it is essential to reduce the noise in PA signals to reconstruct images with less error. In this study, an image reconstruction algorithm for PAM system was implemented and different filtering approaches for denoising were compared. Studies were carried out in three steps: simulation, experimental phantom and blood cell studies. FIR low-pass and band-pass filters and Discrete Wavelet Transform (DWT) based filters (mother wavelets: “bior3.5″, “bior3.7″, “sym7″) with four different thresholding techniques were examined. For the evaluation purposes, Root Mean Square Error (RMSE), Signal to Noise Ratio (SNR) and Contrast to Noise Ratio (CNR) metrics were calculated. In the simulation studies, the most effective methods were obtained as: sym7/heursure/hard thresh. combination (low and medium level noise) and bior3.7/sqtwolog/soft thresh. combination (high-level noise). In experimental phantom studies, noise was classified into five levels. Different filtering approaches perform better depending on the SNR of PA images. For the blood cell study, based on the standard deviation in the background, sym7/sqtwolog/soft thresh. combination provided the best improvement and this result supported the experimental phantom results.

Type: Article
Title: Comparison of noise reduction methods in photoacoustic microscopy
Open access status: An open access version is available from UCL Discovery
DOI: 10.1016/j.compbiomed.2019.04.035
Publisher version: http://dx.doi.org/10.1016/j.compbiomed.2019.04.035
Language: English
Additional information: This version is the author accepted manuscript. For information on re-use, please refer to the publisher’s terms and conditions.
Keywords: Computer Science, Engineering, Photoacoustic (PA), Photoacoustic microscopy (PAM), Discrete wavelet transform (DWT), Signal-to-noise ratio (SNR), Signal denoising, SENSITIVITY
UCL classification: UCL
UCL > Provost and Vice Provost Offices > UCL BEAMS
UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science
UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science > Dept of Med Phys and Biomedical Eng
URI: https://discovery.ucl.ac.uk/id/eprint/10127239
Downloads since deposit
532Downloads
Download activity - last month
Download activity - last 12 months
Downloads by country - last 12 months

Archive Staff Only

View Item View Item