Inter-comparison of relative stopping power estimation models for proton therapy

Doolan, PJ; Collins-Fekete, C-A; Dias, MF; Ruggieri, TA; D'Souza, D; Seco, J; (2016) Inter-comparison of relative stopping power estimation models for proton therapy. Physics in Medicine & Biology , 61 (22) , Article 8085. 10.1088/0031-9155/61/22/8085. Green open access

[thumbnail of Collins FeketeInter-comparison of relative stopping power estimation models for proton therapy_AAM.pdf]
Preview
Text
Collins FeketeInter-comparison of relative stopping power estimation models for proton therapy_AAM.pdf - Accepted Version

Download (2MB) | Preview

Abstract

Theoretical stopping power values were inter-compared for the Bichsel, Janni, ICRU and Schneider relative stopping power (RSP) estimation models, for a variety of tissues and tissue substitute materials taken from the literature. The RSPs of eleven plastic tissue substitutes were measured using Bragg peak shift measurements in water in order to establish a gold standard of RSP values specific to our centre's proton beam characteristics. The theoretical tissue substitute RSP values were computed based on literature compositions to assess the four different computation approaches. The Bichsel/Janni/ICRU approaches led to mean errors in the RSP of  −0.1/+0.7/−0.8%, respectively. Errors when using the Schneider approach, with I-values from the Bichsel, Janni and ICRU sources, followed the same pattern but were generally larger. Following this, the mean elemental ionisation energies were optimized until the differences between theoretical RSP values matched measurements. Failing to use optimized I-values when applying the Schneider technique to 72 human tissues could introduce errors in the RSP of up to  −1.7/+1.1/−0.4% when using Bichsel/Janni/ICRU I-values, respectively. As such, it may be necessary to introduce an additional step in the current stoichiometric calibration procedure in which tissue insert RSPs are measured in a proton beam. Elemental I-values can then optimized to match these measurements, reducing the uncertainty when calculating human tissue RSPs.

Type: Article
Title: Inter-comparison of relative stopping power estimation models for proton therapy
Open access status: An open access version is available from UCL Discovery
DOI: 10.1088/0031-9155/61/22/8085
Publisher version: https://doi.org/10.1088/0031-9155/61/22/8085
Language: English
Additional information: This version is the author accepted manuscript. For information on re-use, please refer to the publisher's terms and conditions.
UCL classification: UCL
UCL > Provost and Vice Provost Offices > UCL BEAMS
UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science
UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science > Dept of Med Phys and Biomedical Eng
URI: https://discovery.ucl.ac.uk/id/eprint/10113775
Downloads since deposit
239Downloads
Download activity - last month
Download activity - last 12 months
Downloads by country - last 12 months

Archive Staff Only

View Item View Item