Plug and play perimetry: Evaluating the use of a self-calibrating digital display for screen-based threshold perimetry

Kyu Han, H; Jones, PR; (2019) Plug and play perimetry: Evaluating the use of a self-calibrating digital display for screen-based threshold perimetry. Displays , 60 pp. 30-38. 10.1016/j.displa.2019.08.006. Green open access

[thumbnail of Jones_DISPLA_2018_176_R1_v2_1.pdf]
Preview
Text
Jones_DISPLA_2018_176_R1_v2_1.pdf - Accepted Version

Download (2MB) | Preview

Abstract

This study evaluated the feasibility of using a ‘self-calibrating’ display (EIZO CG277) to perform screen-based threshold perimetry. Such displays incorporate their own integrated photometer, so could potentially be used ‘straight out of the box’, without the need for time-consuming and costly luminance calibration by skilled experts. Concerns remain, however, due to the fact that the internal calibration of such devices is imperfect (i.e., is limited to a single screen location only) and due to lingering doubts regarding the accuracy of screen-based perimetry in general. To evaluate such a system, automated static threshold perimetry was performed in thirty-two normal-sighted adults. In one condition, participants performed a novel screen-based perimetry test, for which the screen was extensively calibrated using traditional photometric techniques/equipment. In a second condition, the same test was performed, but the display was calibrated using only the screen’s integrated photometer (and assuming uniformity across the display). For reference, participants also completed a traditional visual-field assessment using a Humphrey Field Analyzer (HFA). All three tests were performed twice to assess test-retest repeatability (six tests total). The results showed no differences when comparing screen-based perimetric measurements made with internal self-calibration vs full manual calibration (either in terms of mean sensitivity, pointwise sensitivity, test-retest repeatability, or test duration). Furthermore, the accuracy and precision of both were indistinguishable from the current gold standard (HFA), although the HFA was approximately two minutes (∼30%) faster. These results indicate that self-calibrating commercial monitors can be used to perform screen-based perimetry almost as well as current clinical devices, and without the need for any specialized knowledge or equipment to setup or maintain. This could facilitate perimetric testing in currently hard-to-reach settings, such as community centers, stroke wards, homes, rural locations, or developing countries.

Type: Article
Title: Plug and play perimetry: Evaluating the use of a self-calibrating digital display for screen-based threshold perimetry
Open access status: An open access version is available from UCL Discovery
DOI: 10.1016/j.displa.2019.08.006
Publisher version: https://doi.org/10.1016/j.displa.2019.08.006
Language: English
Additional information: This version is the author accepted manuscript. For information on re-use, please refer to the publisher’s terms and conditions.
Keywords: Visual Fields, Perimetry, Contrast Sensitivity, Screen Calibration, Photometry, Eye-movements
UCL classification: UCL
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Brain Sciences
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Brain Sciences > Institute of Ophthalmology
URI: https://discovery.ucl.ac.uk/id/eprint/10080602
Downloads since deposit
141Downloads
Download activity - last month
Download activity - last 12 months
Downloads by country - last 12 months

Archive Staff Only

View Item View Item