Haemodynamic Issues with Transcatheter Aortic Valve Implantation

Salmonsmith, J; Tango, AM; Ducci, A; Burriesci, G; (2019) Haemodynamic Issues with Transcatheter Aortic Valve Implantation. In: Giordano, A and Biondi-Zoccai, G and Frati, G, (eds.) Transcatheter Aortic Valve Implantation Clinical, Interventional and Surgical Perspectives. (pp. 47-49). Springer: Cham, Switzerland. Green open access

[thumbnail of Burriesci Hemodynamic issues with transcatheter aortic valve implantation[3].pdf]
Preview
Text
Burriesci Hemodynamic issues with transcatheter aortic valve implantation[3].pdf - Accepted Version

Download (444kB) | Preview

Abstract

Transcatheter aortic valves are typically implanted inside the native (or failed bioprosthetic’s) leaflets, permanently forcing the old leaflets open into a pseudo-cylindrical condition. Due to the passive nature of heart valves, the dynamics of the surrounding fluid environment is critical to their optimum performance. Following intervention, the haemodynamics of the region would ideally be returned to their healthy, physiological state, but major alterations are currently inevitable, such as increased peak flow velocity, the presence of stagnation regions, and increased haemolytic fluid environments. These leaflets reduce the volume of and restrict the flow into the Valsalva’s sinuses, and minimise the development of vortices and associated flow structures, which would aid washout and valve closure. Despite these differences to the healthy condition, implantation of these devices offers much improved flow from that of a moderately stenotic valve, with reduced transvalvular systolic pressure drop, peak blood velocity, and shear stress, which normally outweighs the disadvantages highlighted above, especially for high-risk patients.

Type: Book chapter
Title: Haemodynamic Issues with Transcatheter Aortic Valve Implantation
ISBN-13: 978-3-030-05912-5
Open access status: An open access version is available from UCL Discovery
DOI: 10.1007/978-3-030-05912-5_5
Publisher version: https://doi.org/10.1007/978-3-030-05912-5_5
Language: English
Additional information: This version is the author accepted manuscript. For information on re-use, please refer to the publisher’s terms and conditions.
Keywords: In vitro, In simulo, Haemodynamics, Vortex Stagnation, Transcatheter aortic valve
UCL classification: UCL
UCL > Provost and Vice Provost Offices > UCL BEAMS
UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science
UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science > Dept of Mechanical Engineering
URI: https://discovery.ucl.ac.uk/id/eprint/10077489
Downloads since deposit
73Downloads
Download activity - last month
Download activity - last 12 months
Downloads by country - last 12 months

Archive Staff Only

View Item View Item