Benefits of Using a Spatially-variant Penalty Strength with Anatomical Priors in PET Reconstruction

Tsai, Y-J; Schramm, G; Ahn, S; Bousse, A; Arridge, S; Nuyts, J; Hutton, BF; ... Thielemans, K; + view all (2019) Benefits of Using a Spatially-variant Penalty Strength with Anatomical Priors in PET Reconstruction. IEEE Transactions on Medical Imaging 10.1109/TMI.2019.2913889. (In press). Green open access

[thumbnail of 08721278.pdf]
Preview
Text
08721278.pdf - Accepted Version

Download (955kB) | Preview

Abstract

In this study, we explore the use of a spatiallyvariant penalty strength in penalized image reconstruction using anatomical priors to reduce the dependence of lesion contrast on surrounding activity and lesion location. This work builds on a previous method to make the local perturbation response (LPR) approximately spatially invariant. While the dependence of lesion contrast on the local properties introduced by the anatomical penalty is intentional, the method aims to reduce the influence from surroundings lying along the lines of response (LORs) but not in the penalty neighborhood structure. The method is evaluated using simulated data, assuming that the anatomical information is absent or well-aligned with the corresponding activity images. Since the Parallel Level Sets (PLS) penalty is convex and has shown promising results in the literature, it is chosen as the representative anatomical penalty and incorporated into the previously proposed preconditioned algorithm (L-BFGSB-PC) for achieving both good image quality and fast convergence rate. A 2D disc phantom with a feature at the center and a 3D XCAT thorax phantom with lesions inserted in different slices are used to study how surrounding activity and lesion location affect the visual appearance and quantitative consistency. A bias and noise analysis is also performed with the 2D disc phantom. The consistency of the algorithm convergence rate with respect to different data noise and background levels is also investigated using the XCAT phantom. Finally, an example reconstruction for a patient dataset with inserted pseudo lesions is used as a demonstration in a clinical context. We show that applying the spatially-variant penalization with PLS can reduce the dependence of the lesion contrast on the surrounding activity and lesion location. It does not affect the bias and noise trade-off curves for matched local resolution. Moreover, when using the proposed penalization, significant improvement in algorithm convergence rate and convergence consistency is observed.

Type: Article
Title: Benefits of Using a Spatially-variant Penalty Strength with Anatomical Priors in PET Reconstruction
Location: United States
Open access status: An open access version is available from UCL Discovery
DOI: 10.1109/TMI.2019.2913889
Publisher version: https://doi.org/10.1109/TMI.2019.2913889
Language: English
Additional information: This version is the author accepted manuscript. For information on re-use, please refer to the publisher’s terms and conditions.
Keywords: Image reconstruction, Lesion, Perturbation methods, Convergence, Phantoms, Two dimensional displays, Three-dimensional displays
UCL classification: UCL
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Medical Sciences
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Medical Sciences > Div of Medicine
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Medical Sciences > Div of Medicine > Department of Imaging
UCL > Provost and Vice Provost Offices > UCL BEAMS
UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science
UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science > Dept of Computer Science
URI: https://discovery.ucl.ac.uk/id/eprint/10075529
Downloads since deposit
168Downloads
Download activity - last month
Download activity - last 12 months
Downloads by country - last 12 months

Archive Staff Only

View Item View Item