UCL Discovery
UCL home » Library Services » Electronic resources » UCL Discovery

Clinical impact of cardiovascular magnetic resonance with optimized myocardial scar detection in patients with cardiac implantable devices

Bhuva, AN; Kellman, P; Graham, A; Ramlall, M; Boubertakh, R; Feuchter, P; Hawkins, A; ... Manisty, CH; + view all (2019) Clinical impact of cardiovascular magnetic resonance with optimized myocardial scar detection in patients with cardiac implantable devices. International Journal of Cardiology , 279 pp. 72-78. 10.1016/j.ijcard.2019.01.005. Green open access

[thumbnail of Bhuva_Wideband_13122018.pdf]
Preview
Text
Bhuva_Wideband_13122018.pdf - Accepted Version

Download (444kB) | Preview

Abstract

BACKGROUND: Myocardial scar assessment using late gadolinium enhancement Cardiovascular Magnetic Resonance (LGE CMR) is commonly indicated for patients with cardiac implantable electronic devices (CIEDs), however metal artifact can degrade images. We evaluated the clinical impact of LGE CMR incorporating a device-dependent metal artifact reduction strategy in patients with CIEDs. METHODS: 136 CMR studies were performed in 133 consecutive patients (age 56 ± 19 years, 69% male) with CIEDs (22% implantable loop recorders [ILRs], 40% permanent pacemakers [PPMs], 38% implantable cardioverter defibrillators [ICDs]; 42% non-MRI conditional) over 2 years, without complication. LGE imaging was tailored to the CIED, using a wideband sequence for left-sided PPMs and ICDs and conventional sequences for ILRs and right-sided PPMs, scoring segmental artifact. Diagnostic utility and impact on clinical management were scored by consensus of experts. RESULTS: CMR provided unexpected diagnoses in 22 (16%) and changed management in 113 (83%) patients. Myocardial scar was present in 92 (68%), with other abnormalities detected in another 13%. Using conventional LGE, 43 (32%) studies were non-diagnostic (79% of defibrillators) compared to 0% using wideband LGE imaging. Wideband LGE results changed clinical management in an additional 39 (75%) defibrillator patients and 10 (19%) pacemaker patients when compared to imaging with conventional LGE sequences. CONCLUSION: The clinical yield from CMR using optimized LGE sequences in patients with CIEDs is high with no demonstrated clinical risk. A device-dependent LGE imaging strategy using wideband LGE is needed to achieve clinical utility especially in ICD recipients.

Type: Article
Title: Clinical impact of cardiovascular magnetic resonance with optimized myocardial scar detection in patients with cardiac implantable devices
Location: Netherlands
Open access status: An open access version is available from UCL Discovery
DOI: 10.1016/j.ijcard.2019.01.005
Publisher version: http://doi.org/10.1016/j.ijcard.2019.01.005
Language: English
Additional information: This version is the author accepted manuscript. For information on re-use, please refer to the publisher’s terms and conditions.
Keywords: Cardiovascular magnetic resonance, Defibrillator, Fibrosis, Late gadolinium enhancement, Pacemaker, Wideband
UCL classification: UCL
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Population Health Sciences > Institute of Cardiovascular Science
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Population Health Sciences > Institute of Cardiovascular Science > Clinical Science
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Population Health Sciences > Institute of Health Informatics
URI: https://discovery.ucl.ac.uk/id/eprint/10067494
Downloads since deposit
332Downloads
Download activity - last month
Download activity - last 12 months
Downloads by country - last 12 months

Archive Staff Only

View Item View Item