UCL Discovery
UCL home » Library Services » Electronic resources » UCL Discovery

Impact of varying planning parameters on proton pencil beam scanning dose distributions in four commercial treatment planning systems

Alshaikhi, J; Doolan, PJ; D'Souza, D; Holloway, SM; Amos, RA; Royle, G; (2019) Impact of varying planning parameters on proton pencil beam scanning dose distributions in four commercial treatment planning systems. Medical Physics , 46 (3) pp. 1150-1162. 10.1002/mp.13382. Green open access

[thumbnail of Alshaikhi_final_5Feb19.pdf]
Preview
Text
Alshaikhi_final_5Feb19.pdf - Accepted Version

Download (2MB) | Preview

Abstract

PURPOSE: In pencil beam scanning proton therapy, target coverage is achieved by scanning the pencil beam laterally in the x- and y-directions and delivering spots of dose to positions at a given radiological depth (layer). Dose is delivered to the spots on different layers by pencil beams of different energy until the entire volume has been irradiated. The aim of this study is to investigate the implementation of proton planning parameters (spot spacing, layer spacing and margins) in four commercial proton treatment planning systems (TPSs): Eclipse, Pinnacle3 , RayStation and XiO. MATERIALS AND METHODS: Using identical beam data in each TPS, plans were created on uniform material synthetic phantoms with cubic targets. The following parameters were systematically varied in each TPS to observe their different implementations: spot spacing, layer spacing and margin. Additionally, plans were created in Eclipse to investigate the impact of these parameters on plan delivery and optimal values are suggested. RESULTS: It was found that all systems except Eclipse use a variable layer spacing per beam, based on the Bragg peak width of each energy layer. It is recommended that if this cannot be used, then a constant value of 5 mm will ensure good dose homogeneity. Only RayStation varies the spot spacing according to the variable spot size with depth. If a constant spot spacing is to be used, a value of 5 mm is recommended as a good compromise between dose homogeneity, plan robustness and planning time. It was found that both Pinnacle3 and RayStation position spots outside of the defined volume (target plus margin). CONCLUSIONS: All four systems are capable of delivering uniform dose distributions to simple targets, but their implementation of the various planning parameters is different. In this paper comparisons are made between the four systems and recommendations are made as to the values that will provide the best compromise in dose homogeneity and planning time. This article is protected by copyright. All rights reserved.

Type: Article
Title: Impact of varying planning parameters on proton pencil beam scanning dose distributions in four commercial treatment planning systems
Location: United States
Open access status: An open access version is available from UCL Discovery
DOI: 10.1002/mp.13382
Publisher version: https://doi.org/10.1002/mp.13382
Language: English
Additional information: This version is the author accepted manuscript. For information on re-use, please refer to the publisher’s terms and conditions.
Keywords: Proton therapy, particle therapy, treatment planning
UCL classification: UCL
UCL > Provost and Vice Provost Offices > UCL BEAMS
UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science
UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science > Dept of Med Phys and Biomedical Eng
URI: https://discovery.ucl.ac.uk/id/eprint/10067010
Downloads since deposit