UCL Discovery
UCL home » Library Services » Electronic resources » UCL Discovery

Retrieval Findings of Recalled Dual-Taper Hips

Di Laura, A; Hothi, HS; Henckel, J; Kwon, Y-M; Skinner, JA; Hart, AJ; (2018) Retrieval Findings of Recalled Dual-Taper Hips. The Journal of Bone and Joint Surgery , 100 (19) pp. 1661-1672. 10.2106/JBJS.17.00790. Green open access

[thumbnail of Retrieval_Findings_of_Recalled_Dual_Taper_Hips.5.pdf]
Preview
Text
Retrieval_Findings_of_Recalled_Dual_Taper_Hips.5.pdf - Published Version

Download (3MB) | Preview

Abstract

BACKGROUND: The recent high-profile recalls of several dual-taper hip designs pose questions regarding why those designs perform poorly. We aimed to characterize taper damage in 1 recalled design to understand failure mechanisms to inform surgeons on which patients should be considered at risk of revision and when to revise. METHODS: High-precision measurement equipment was used to characterize the metal loss from the neck-stem interface of 116 retrieved Rejuvenate femoral stems (Stryker Howmedica Osteonics) revised because of an adverse reaction to metal debris. Head-neck taper surfaces were also investigated, and clinical and laboratory data were examined. RESULTS: The neck-stem junction of each implant was moderately to severely corroded and showed a characteristic wear pattern on both male and female taper surfaces. The severity of taper damage was positively correlated with time to revision (coefficient, 0.040 [95% confidence interval (CI), 0.028 to 0.051]; p < 0.0001) and with serum cobalt concentration (coefficient, 0.02 [95% CI, 0.01 to 0.02]; p < 0.0001) and serum chromium concentration (coefficient, 0.04 [95% CI, 0.009 to 0.070]; p = 0.0142). CONCLUSIONS: A forensic examination of the retrieved components that failed secondary to an adverse reaction to metal debris showed, in all cases, visible corrosion. Of the implant and patient factors investigated, we did not identify any predictors of corrosion. The severity of damage was found to increase with time; cobalt was significantly elevated over chromium. CLINICAL RELEVANCE: Surgeons should scrupulously follow and consider revision for patients with this implant design to avoid extensive tissue excision. Blood metal ion tests may aid in identifying the mechanisms of taper corrosion.

Type: Article
Title: Retrieval Findings of Recalled Dual-Taper Hips
Location: United States
Open access status: An open access version is available from UCL Discovery
DOI: 10.2106/JBJS.17.00790
Publisher version: https://doi.org/10.2106/JBJS.17.00790
Language: English
Additional information: This is the published version of record. For information on re-use, please refer to the publisher’s terms and conditions.
UCL classification: UCL
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Medical Sciences
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Medical Sciences > Div of Surgery and Interventional Sci
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Medical Sciences > Div of Surgery and Interventional Sci > Department of Ortho and MSK Science
UCL > Provost and Vice Provost Offices > UCL BEAMS
UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science
UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science > Dept of Mechanical Engineering
URI: https://discovery.ucl.ac.uk/id/eprint/10058821
Downloads since deposit
127Downloads
Download activity - last month
Download activity - last 12 months
Downloads by country - last 12 months

Archive Staff Only

View Item View Item