UCL Discovery
UCL home » Library Services » Electronic resources » UCL Discovery

On the feasibility of automated mechanical ventilation control through EIT

Tregidgo, H; Crabb, MG; Hazel, A; Lionheart, W; (2018) On the feasibility of automated mechanical ventilation control through EIT. IEEE Transactions on Biomedical Engineering , 65 (11) 2459 -2470. 10.1109/TBME.2018.2798812. Green open access

[thumbnail of Tregidgo VoR 08301569.pdf]
Preview
Text
Tregidgo VoR 08301569.pdf - Published Version

Download (1MB) | Preview

Abstract

OBJECTIVE: This paper aims to demonstrate the feasibility of coupling electrical impedance tomography (EIT) with models of lung function in order to recover parameters and inform mechanical ventilation control. METHODS: A compartmental ordinary differential equation model of lung function is coupled to simulations of EIT, assuming accurate modelling and movement tracking, to generate time series values of bulk conductivity. These values are differentiated and normalised against the total air volume flux to recover regional volumes and flows. These ventilation distributions are used to recover regional resistance and elastance properties of the lung. Linear control theory is used to demonstrate how these parameters may be used to generate a patient-specific pressure mode control. RESULTS: Ventilation distributions are shown to be recoverable, with Euclidean norm errors in air flow below 9% and volume below 3%. The parameters are also shown to be recoverable, although errors are higher for resistance values than elastance. The control constructed is shown to have minimal H¹ semi-norm resulting in bounded magnitudes and minimal gradients. CONCLUSION: The recovery of regional ventilation distributions and lung parameters is feasible with the use of EIT. These parameters may then be used in model based control schemes to provide patient-specific care. SIGNIFICANCE: For pulmonary-intensive-care patients mechanical ventilation is a life saving intervention, careful calibration of pressure settings. Both magnitudes and gradients of pressure can all contribute to ventilator induced lung injury. Retrieving regional lung parameters allows the design of patient-specific ventilator controls to reduce injury.

Type: Article
Title: On the feasibility of automated mechanical ventilation control through EIT
Open access status: An open access version is available from UCL Discovery
DOI: 10.1109/TBME.2018.2798812
Publisher version: https://doi.org/10.1109/TBME.2018.2798812
Language: English
Additional information: © 2018 IEEE. This work is licensed under a Creative Commons Attribution 3.0 License (http://creativecommons.org/licenses/by/3.0/).
Keywords: Lung, Ventilation, Tomography, Mathematical model, Atmospheric modeling, Pressure control, Conductivity
UCL classification: UCL
UCL > Provost and Vice Provost Offices > UCL BEAMS
UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science
UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science > Dept of Med Phys and Biomedical Eng
URI: https://discovery.ucl.ac.uk/id/eprint/10056679
Downloads since deposit
170Downloads
Download activity - last month
Download activity - last 12 months
Downloads by country - last 12 months

Archive Staff Only

View Item View Item