Stergiou, GS;
Alpert, B;
Mieke, S;
Asmar, R;
Atkins, N;
Eckert, S;
Frick, G;
... O'Brien, E; + view all
(2018)
A universal standard for the validation of blood pressure measuring devices: Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Collaboration Statement.
Journal of Hypertension
, 36
(3)
pp. 472-478.
10.1097/HJH.0000000000001634.
Preview |
Text
AAMI-ESH-ISO Statement REVISED 20171006.pdf - Accepted Version Download (716kB) | Preview |
Abstract
In the last 30 years, several organizations, such as the US Association for the Advancement of Medical Instrumentation (AAMI), the British Hypertension Society, the European Society of Hypertension (ESH) Working Group on Blood Pressure (BP) Monitoring and the International Organization for Standardization (ISO) have developed protocols for clinical validation of BP measuring devices. However, it is recognized that science, as well as patients, consumers and manufacturers would be best served if all BP measuring devices were assessed for accuracy according to an agreed single validation protocol that had global acceptance. Therefore, an international initiative was taken by AAMI, ESH and ISO experts who agreed to develop a universal standard for device validation. This statement presents the key aspects of a validation procedure, which were agreed by the AAMI, ESH and ISO representatives as the basis for a single universal validation protocol. As soon as the AAMI/ESH/ISO standard is fully developed, this will be regarded as the single universal standard and will replace all other previous standards/protocols.
Type: | Article |
---|---|
Title: | A universal standard for the validation of blood pressure measuring devices: Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Collaboration Statement |
Open access status: | An open access version is available from UCL Discovery |
DOI: | 10.1097/HJH.0000000000001634 |
Publisher version: | http://dx.doi.org/10.1097/HJH.0000000000001634 |
Language: | English |
Additional information: | This version is the author accepted manuscript. For information on re-use, please refer to the publisher’s terms and conditions. |
Keywords: | Science & Technology, Life Sciences & Biomedicine, Peripheral Vascular Disease, Cardiovascular System & Cardiology, accuracy, blood pressure measurement, device, monitor, protocol, standard, validation, Hypertension-International-Protocol, European-Society, Monitors, Revision, Accuracy, Adults |
UCL classification: | UCL UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Population Health Sciences > Institute of Cardiovascular Science UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Population Health Sciences > Institute of Cardiovascular Science > Population Science and Experimental Medicine |
URI: | https://discovery.ucl.ac.uk/id/eprint/10048123 |
Archive Staff Only