Effect of Bearing Type on Taper Material Loss in Hips From 1 Manufacturer

Hothi, HS; Eskelinen, AP; Henckel, J; Kwon, Y-M; Blunn, GW; Skinner, JA; Hart, AJ; (2018) Effect of Bearing Type on Taper Material Loss in Hips From 1 Manufacturer. The Journal of Arthroplasty , 33 (5) pp. 1588-1593. 10.1016/j.arth.2017.12.022. Green open access

[thumbnail of Manuscript.pdf]
Preview
Text
Manuscript.pdf - Accepted Version

Download (4MB) | Preview

Abstract

BACKGROUND: Numerous studies have reported on clinical significant volumes of material loss and corrosion at the head-stem junction of metal-on-metal (MOM) hips; less is understood about metal-on-polyethylene (MOP) hips. We compared the effect of bearing type (MOM vs MOP) on taper material loss for a hip system of a single design. METHODS: In this cohort study, we recruited retrieved MOM (n = 30) and MOP (n = 22) bearing hips that were consecutively received at our center. We prospectively collected associated clinical and imaging data. We measured the severity of corrosion and volumes of material loss at each head taper surface and used multivariate statistical analysis to investigate differences between the 2 bearing types. RESULTS: The median rate of material loss for the MOM and MOP groups was 0.81 mm3/y (0.01-3.45) and 0.03 mm3/y (0-1.07), respectively (P < .001). Twenty-nine of 30 MOM hips were revised for adverse metal reactions, compared with 1 of 22 MOP hips. CONCLUSION: MOP hips lost significantly less material from their taper junctions than MOM hips. Our results can reassure patients with MOP Pinnacle hips that they are unlikely to experience clinically significant problems related to material loss from the taper junction.

Type: Article
Title: Effect of Bearing Type on Taper Material Loss in Hips From 1 Manufacturer
Location: United States
Open access status: An open access version is available from UCL Discovery
DOI: 10.1016/j.arth.2017.12.022
Publisher version: http://dx.doi.org/10.1016/j.arth.2017.12.022
Language: English
Additional information: This version is the author accepted manuscript. For information on re-use, please refer to the publisher’s terms and conditions.
Keywords: corrosion, hip, metal-on-polyethylene, taper, wear
UCL classification: UCL
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Medical Sciences
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Medical Sciences > Div of Surgery and Interventional Sci
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Medical Sciences > Div of Surgery and Interventional Sci > Department of Ortho and MSK Science
URI: https://discovery.ucl.ac.uk/id/eprint/10046747
Downloads since deposit
96Downloads
Download activity - last month
Download activity - last 12 months
Downloads by country - last 12 months

Archive Staff Only

View Item View Item