Technical Note: Error metrics for estimating the accuracy of needle/instrument placement during transperineal MR/US-guided prostate interventions

Bonmati, E; Hu, Y; Villarini, B; Rodell, R; Martin, P; Han, L; Donaldson, I; ... Barratt, DC; + view all (2018) Technical Note: Error metrics for estimating the accuracy of needle/instrument placement during transperineal MR/US-guided prostate interventions. Medical Physics 10.1002/mp.12814. (In press). Green open access

[thumbnail of Bonmati_Technical_Note_Error_metrics_AAM.pdf]
Preview
Text
Bonmati_Technical_Note_Error_metrics_AAM.pdf - Accepted Version

Download (271kB) | Preview

Abstract

PURPOSE: Image-guided systems that fuse magnetic resonance imaging (MRI) with three-dimensional (3D) ultrasound (US) images for performing targeted prostate needle biopsy and minimally-invasive treatments for prostate cancer are of increasing clinical interest. To date, a wide range of different accuracy estimation procedures and error metrics have been reported, which makes comparing the performance of different systems difficult. METHODS: A set of 9 measures are presented to assess the accuracy of MRI-US image registration, needle positioning, needle guidance, and overall system error, with the aim of providing a methodology for estimating the accuracy of instrument placement using a MR/US-guided transperineal approach. RESULTS: Using the SmartTarget fusion system, an MRI-US image alignment error was determined to be 2.0±1.0 mm (mean ± SD), and an overall system instrument targeting error of 3.0±1.2 mm. Three needle deployments for each target phantom lesion was found to result in a 100% lesion hit rate and a median predicted cancer core length of 5.2 mm. CONCLUSIONS: The application of a comprehensive, unbiased validation assessment for MR/TRUS guided systems can provide useful information on system performance for quality assurance and system comparison. Furthermore, such an analysis can be helpful in identifying relationships between these errors, providing insight into the technical behaviour of these systems.

Type: Article
Title: Technical Note: Error metrics for estimating the accuracy of needle/instrument placement during transperineal MR/US-guided prostate interventions
Open access status: An open access version is available from UCL Discovery
DOI: 10.1002/mp.12814
Publisher version: http://dx.doi.org/10.1002/mp.12814
Language: English
Additional information: This version is the author accepted manuscript. For information on re-use, please refer to the publisher’s terms and conditions.
Keywords: Accuracy Validation; Needle Placement; Prostate Cancer; Targeted Biopsy; Focal Therapy; Image-guided Interventions
UCL classification: UCL
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Medical Sciences
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Medical Sciences > Div of Surgery and Interventional Sci
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Medical Sciences > Div of Surgery and Interventional Sci > Department of Targeted Intervention
UCL > Provost and Vice Provost Offices > UCL BEAMS
UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science
UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science > Dept of Med Phys and Biomedical Eng
URI: https://discovery.ucl.ac.uk/id/eprint/10043388
Downloads since deposit
125Downloads
Download activity - last month
Download activity - last 12 months
Downloads by country - last 12 months

Archive Staff Only

View Item View Item